Your browser doesn't support javascript.
loading
Vein visualization: patient characteristic factors and efficacy of a new infrared vein finder technology.
Chiao, F B; Resta-Flarer, F; Lesser, J; Ng, J; Ganz, A; Pino-Luey, D; Bennett, H; Perkins, C; Witek, B.
Afiliación
  • Chiao FB; Department of Anesthesiology, St Lukes Roosevelt Hospital Center, Columbia University College of Physicians and Surgeons, New York, NY, USA. fchiao@gmail.com
Br J Anaesth ; 110(6): 966-71, 2013 Jun.
Article en En | MEDLINE | ID: mdl-23384732
ABSTRACT

BACKGROUND:

We investigated the patient characteristic factors that correlate with identification of i.v. cannulation sites with normal eyesight. We evaluated a new infrared vein finding (VF) technology device in identifying i.v. cannulation sites.

METHODS:

Each subject underwent two observations one using the conventional method (CM) of normal, unassisted eyesight and the other with the infrared VF device, VueTek's Veinsite™ (VF). A power analysis for moderate effect size (ß=0.95) required 54 samples for within-subject differences.

RESULTS:

Patient characteristic profiles were obtained from 384 subjects (768 observations). Our sample population exhibited an overall average of 5.8 [95% confidence interval (CI) 5.4-6.2] veins using CM. As a whole, CM vein visualization were less effective among obese [4.5 (95% CI 3.8-5.3)], African-American [4.6 (95% CI 3.6-5.5 veins)], and Asian [5.1 (95% CI 4.1-6.0)] subjects. Next, the VF technology identified an average of 9.1 (95% CI 8.6-9.5) possible cannulation sites compared with CM [average of 5.8 (95% CI 5.4-6.2)]. Seventy-six obese subjects had an average of 4.5 (95% CI 3.8-5.3) and 8.2 (95% CI 7.4-9.1) veins viewable by CM and VF, respectively. In dark skin subjects, 9.1 (95% CI 8.3-9.9) veins were visible by VF compared with 5.4 (95% CI 4.8-6.0) with CM.

CONCLUSIONS:

African-American or Asian ethnicity, and obesity were associated with decreased vein visibility. The visibility of veins eligible for cannulation increased for all subgroups using a new infrared device.
Asunto(s)
Palabras clave

Texto completo: 1 Colección: 01-internacional Banco de datos: MEDLINE Asunto principal: Venas / Cateterismo / Infarto Tipo de estudio: Prognostic_studies Límite: Adolescent / Adult / Aged / Child / Child, preschool / Female / Humans / Infant / Male / Middle aged Idioma: En Revista: Br J Anaesth Año: 2013 Tipo del documento: Article País de afiliación: Estados Unidos

Texto completo: 1 Colección: 01-internacional Banco de datos: MEDLINE Asunto principal: Venas / Cateterismo / Infarto Tipo de estudio: Prognostic_studies Límite: Adolescent / Adult / Aged / Child / Child, preschool / Female / Humans / Infant / Male / Middle aged Idioma: En Revista: Br J Anaesth Año: 2013 Tipo del documento: Article País de afiliación: Estados Unidos