Bench test assessment of mainstream capnography during high frequency oscillatory ventilation.
Hartdorff, Caroline M; van Heerde, Marc; Markhorst, Dick G.
Afiliación
  • Hartdorff CM; Pediatric Intensive Care Unit, VU University Medical Center, PO BOX 7057, 1007 MB, Amsterdam, The Netherlands.
J Clin Monit Comput ; 28(1): 63-6, 2014 Feb.
Article en En | MEDLINE | ID: mdl-23974630
ABSTRACT
To assess the feasibility, stability and predictability of pCO2 measurement (PETCO2) using a main stream capnograph in a high frequency oscillatory ventilation circuit. A commercially available capnograph was mounted into a high frequency oscillatory ventilator patient circuit, adjustable CO2 flow was introduced into an artificial lung and the output of the CO2 sensor assessed under varying ventilator settings. Influence of oxygen content, pressures, heat and moisture were recorded. A linear relationship between CO2 flow rate and PETCO2 was found. Varying ventilator settings influenced the measurements, but the results for PETCO2 remained within a range of 1.5 mmHg above or under then mean measurement value. Measurements remained stable despite humidification, heat, pressure amplitudes or mean airway pressure changes. From this bench test, we conclude it is feasible to measure PETCO2 using a main stream capnograph during high frequency oscillatory conditions, these measurements were stable during the experiment. Changes in CO2 production or output can be detected. The system may prove to be of clinical value, but further in vivo measurements are warranted.
Asunto(s)

Texto completo: 1 Colección: 01-internacional Banco de datos: MEDLINE Asunto principal: Oscilometría / Respiración Artificial / Ventilación de Alta Frecuencia / Capnografía Tipo de estudio: Prognostic_studies Límite: Humans Idioma: En Revista: J Clin Monit Comput Asunto de la revista: INFORMATICA MEDICA / MEDICINA Año: 2014 Tipo del documento: Article País de afiliación: Países Bajos

Texto completo: 1 Colección: 01-internacional Banco de datos: MEDLINE Asunto principal: Oscilometría / Respiración Artificial / Ventilación de Alta Frecuencia / Capnografía Tipo de estudio: Prognostic_studies Límite: Humans Idioma: En Revista: J Clin Monit Comput Asunto de la revista: INFORMATICA MEDICA / MEDICINA Año: 2014 Tipo del documento: Article País de afiliación: Países Bajos