Pulsed-light imaging for fluorescence guided surgery under normal room lighting.
Sexton, Kristian; Davis, Scott C; McClatchy, David; Valdes, Pablo A; Kanick, Stephen C; Paulsen, Keith D; Roberts, David W; Pogue, Brian W.
Afiliación
  • Sexton K; Thayer School of Engineering at Dartmouth College, Hanover, New Hampshire 03755, USA. kristian.j.sexton.th@dartmouth.edu
Opt Lett ; 38(17): 3249-52, 2013 Sep 01.
Article en En | MEDLINE | ID: mdl-23988926
ABSTRACT
Fluorescence guided surgery (FGS) is an emerging technology that has demonstrated improved surgical outcomes. However, dim lighting conditions required by current FGS systems are disruptive to standard surgical workflow. We present a novel FGS system capable of imaging fluorescence under normal room light by using pulsed excitation and gated acquisition. Images from tissue-simulating phantoms confirm visual detection down to 0.25 µM of protoporphyrin IX under 125 µW/cm2 of ambient light, more than an order of magnitude lower than that measured with the Zeiss Pentero in the dark. Resection of orthotopic brain tumors in mice also suggests that the pulsed-light system provides superior sensitivity in vivo.
Asunto(s)

Texto completo: 1 Colección: 01-internacional Banco de datos: MEDLINE Asunto principal: Iluminación / Cirugía Asistida por Computador / Fluorescencia / Luz Límite: Animals Idioma: En Revista: Opt Lett Año: 2013 Tipo del documento: Article País de afiliación: Estados Unidos

Texto completo: 1 Colección: 01-internacional Banco de datos: MEDLINE Asunto principal: Iluminación / Cirugía Asistida por Computador / Fluorescencia / Luz Límite: Animals Idioma: En Revista: Opt Lett Año: 2013 Tipo del documento: Article País de afiliación: Estados Unidos