Liquid chromatography high-resolution TOF analysis: investigation of MSE for broad-spectrum drug screening.
Chindarkar, Nandkishor S; Wakefield, Michael R; Stone, Judith A; Fitzgerald, Robert L.
Afiliación
  • Chindarkar NS; Department of Pathology, Center for Advanced Laboratory Medicine, University of California, San Diego Health Systems, San Diego, CA; nchindarkar@ucsd.edu.
  • Wakefield MR; Waters Corp., Pleasanton, CA.
  • Stone JA; Department of Pathology, Center for Advanced Laboratory Medicine, University of California, San Diego Health Systems, San Diego, CA;
  • Fitzgerald RL; Department of Pathology, Center for Advanced Laboratory Medicine, University of California, San Diego Health Systems, San Diego, CA;
Clin Chem ; 60(8): 1115-25, 2014 Aug.
Article en En | MEDLINE | ID: mdl-24916795
ABSTRACT

BACKGROUND:

High-resolution mass spectrometry (HRMS) has the potential to supplement other drug screening platforms used in toxicology laboratories. HRMS offers high analytical specificity, which can be further enhanced by incorporating a fragment ion for each analyte. The ability to obtain precursor ions and fragment ions using elevated collision energies (MS(E)) can help improve the specificity of HRMS methods.

METHODS:

We developed a broad-spectrum screening method on an ultraperformance liquid chromatography TOF mass spectrometer (UPLC-TOF-MS) using the MS(E) mode. A diverse set of patient samples were subjected to a simple dilute, hydrolyze, and shoot protocol and analyzed in a blind manner. Data were processed with 3 sets of criteria with increasing stringency, and the results were compared with the reference laboratory results.

RESULTS:

A combination of retention time match (±0.2 min), a protonated analyte, and fragment ion mass accuracy of ±5 ppm produced zero false-positive results. Using these criteria, we confirmed 92% (253/275) of true positives. The positive confirmation rate increased to 98% (270/275) when the requirement for a fragment ion was dropped, but also produced 53 false positives. A total of 136 additional positive drug findings not identified by the reference methods were identified with the UPLC-TOF-MS.

CONCLUSIONS:

MS(E) provides a unique way to incorporate fragment ion information without the need of precursor ion selection. A primary limitation of requiring a fragment ion for positive identification was that certain drug classes required high-energy collisions, which formed many fragment ions of low abundance that were not readily detected.
Asunto(s)

Texto completo: 1 Colección: 01-internacional Banco de datos: MEDLINE Asunto principal: Espectrometría de Masas / Cromatografía Liquida / Monitoreo de Drogas Tipo de estudio: Clinical_trials / Diagnostic_studies / Prognostic_studies / Screening_studies Límite: Humans Idioma: En Revista: Clin Chem Asunto de la revista: QUIMICA CLINICA Año: 2014 Tipo del documento: Article

Texto completo: 1 Colección: 01-internacional Banco de datos: MEDLINE Asunto principal: Espectrometría de Masas / Cromatografía Liquida / Monitoreo de Drogas Tipo de estudio: Clinical_trials / Diagnostic_studies / Prognostic_studies / Screening_studies Límite: Humans Idioma: En Revista: Clin Chem Asunto de la revista: QUIMICA CLINICA Año: 2014 Tipo del documento: Article