Quantitative high-efficiency cadmium-zinc-telluride SPECT with dedicated parallel-hole collimation system in obese patients: results of a multi-center study.
Nakazato, Ryo; Slomka, Piotr J; Fish, Mathews; Schwartz, Ronald G; Hayes, Sean W; Thomson, Louise E J; Friedman, John D; Lemley, Mark; Mackin, Maria L; Peterson, Benjamin; Schwartz, Arielle M; Doran, Jesse A; Germano, Guido; Berman, Daniel S.
Afiliación
  • Nakazato R; Department of Imaging, Cedars-Sinai Medical Center, 8700 Beverly Boulevard, Taper Building Suite 1258, Los Angeles, CA, 90048, USA.
J Nucl Cardiol ; 22(2): 266-75, 2015 Apr.
Article en En | MEDLINE | ID: mdl-25388380
ABSTRACT

BACKGROUND:

Obesity is a common source of artifact on conventional SPECT myocardial perfusion imaging (MPI). We evaluated image quality and diagnostic performance of high-efficiency (HE) cadmium-zinc-telluride parallel-hole SPECT MPI for coronary artery disease (CAD) in obese patients. METHODS AND

RESULTS:

118 consecutive obese patients at three centers (BMI 43.6 ± 8.9 kg·m(-2), range 35-79.7 kg·m(-2)) had upright/supine HE-SPECT and invasive coronary angiography > 6 months (n = 67) or low likelihood of CAD (n = 51). Stress quantitative total perfusion deficit (TPD) for upright (U-TPD), supine (S-TPD), and combined acquisitions (C-TPD) was assessed. Image quality (IQ; 5 = excellent; < 3 nondiagnostic) was compared among BMI 35-39.9 (n = 58), 40-44.9 (n = 24) and ≥45 (n = 36) groups. ROC curve area for CAD detection (≥50% stenosis) for U-TPD, S-TPD, and C-TPD were 0.80, 0.80, and 0.87, respectively. Sensitivity/specificity was 82%/57% for U-TPD, 74%/71% for S-TPD, and 80%/82% for C-TPD. C-TPD had highest specificity (P = .02). C-TPD normalcy rate was higher than U-TPD (88% vs 75%, P = .02). Mean IQ was similar among BMI 35-39.9, 40-44.9 and ≥45 groups [4.6 vs 4.4 vs 4.5, respectively (P = .6)]. No patient had a nondiagnostic stress scan.

CONCLUSIONS:

In obese patients, HE-SPECT MPI with dedicated parallel-hole collimation demonstrated high image quality, normalcy rate, and diagnostic accuracy for CAD by quantitative analysis of combined upright/supine acquisitions.
Asunto(s)

Texto completo: 1 Colección: 01-internacional Banco de datos: MEDLINE Asunto principal: Enfermedad de la Arteria Coronaria / Aumento de la Imagen / Tomografía Computarizada de Emisión de Fotón Único / Artefactos Tipo de estudio: Clinical_trials / Diagnostic_studies / Evaluation_studies / Prognostic_studies Límite: Female / Humans / Male / Middle aged País/Región como asunto: America do norte Idioma: En Revista: J Nucl Cardiol Asunto de la revista: CARDIOLOGIA Año: 2015 Tipo del documento: Article País de afiliación: Estados Unidos

Texto completo: 1 Colección: 01-internacional Banco de datos: MEDLINE Asunto principal: Enfermedad de la Arteria Coronaria / Aumento de la Imagen / Tomografía Computarizada de Emisión de Fotón Único / Artefactos Tipo de estudio: Clinical_trials / Diagnostic_studies / Evaluation_studies / Prognostic_studies Límite: Female / Humans / Male / Middle aged País/Región como asunto: America do norte Idioma: En Revista: J Nucl Cardiol Asunto de la revista: CARDIOLOGIA Año: 2015 Tipo del documento: Article País de afiliación: Estados Unidos