A comparative study of strain and shear-wave elastography in an elasticity phantom.
Carlsen, Jonathan F; Pedersen, Malene R; Ewertsen, Caroline; Saftoiu, Adrian; Lönn, Lars; Rafaelsen, Søren R; Nielsen, Michael B.
Afiliación
  • Carlsen JF; 1 Department of Radiology, Rigshospitalet, Copenhagen University Hospital, Blegdamsvej 9, 2100 Copenhagen, Denmark.
AJR Am J Roentgenol ; 204(3): W236-42, 2015 Mar.
Article en En | MEDLINE | ID: mdl-25714307
ABSTRACT
OBJECTIVE. The purpose of this study was to assess the diagnostic accuracy of strain and shear-wave elastography for determining targets of varying stiffness in a phantom. The effect of target diameter on elastographic assessments and the effect of depth on shear-wave velocity were also investigated. MATERIALS AND METHODS. We examined 20 targets of varying diameters (2.5-16.7 mm) and stiffnesses (8, 14, 45, and 80 kPa) with a 4-9-MHz linear-array transducer. Targets were evaluated 10 times with three different methods-shear-wave elastography, strain ratio, and strain histogram analysis-yielding 600 evaluations. AUCs were calculated for data divided between different stiffnesses. A 1.5-6-MHz curved-array transducer was used to assess the effect of depth (3.5 vs 6 cm) on shear-wave elastography in 80 scans. Mixed model analysis was performed to assess the effect of target diameter and depth. RESULTS. Strain ratio and strain histogram AUCs were higher than the shear-wave velocity AUC (p < 0.001) in data divided as 80 versus 45, 14, and 8 kPa. In data divided as 80 and 45 versus 14 and 8 kPa, the methods were equal (p = 0.959 and p = 1.000, respectively). Strain ratios were superior (p = 0.030), whereas strain histograms were not significantly better (p = 0.083) than shear-wave elastography in data divided as 80, 45, and 14 versus 8 kPa. Target diameter had an effect on all three methods (p = 0.001). Depth had an effect on shear-wave velocity (p = 0.001). CONCLUSION. The ability to discern different target stiffnesses varies between shear-wave and strain elastography. Target diameter affected all methods. Shear-wave elastography is affected by target depth.
Asunto(s)
Palabras clave

Texto completo: 1 Colección: 01-internacional Banco de datos: MEDLINE Asunto principal: Fantasmas de Imagen / Diagnóstico por Imagen de Elasticidad Tipo de estudio: Diagnostic_studies / Evaluation_studies / Prognostic_studies Idioma: En Revista: AJR Am J Roentgenol Año: 2015 Tipo del documento: Article País de afiliación: Dinamarca

Texto completo: 1 Colección: 01-internacional Banco de datos: MEDLINE Asunto principal: Fantasmas de Imagen / Diagnóstico por Imagen de Elasticidad Tipo de estudio: Diagnostic_studies / Evaluation_studies / Prognostic_studies Idioma: En Revista: AJR Am J Roentgenol Año: 2015 Tipo del documento: Article País de afiliación: Dinamarca