Your browser doesn't support javascript.
loading
Aptamer-based detection and quantitative analysis of human immunoglobulin E in capillary electrophoresis with chemiluminescence detection.
Liu, Yan-Ming; Cao, Jun-Tao; Liu, Ying-Ying; Zhang, Jing-Jing; Zhou, Min; Huang, Ke-Jing; Chen, Yong-Hong; Ren, Shu-Wei.
Afiliación
  • Liu YM; College of Chemistry and Chemical Engineering, Xinyang Normal University, Xinyang, P. R. China.
  • Cao JT; College of Chemistry and Chemical Engineering, Xinyang Normal University, Xinyang, P. R. China.
  • Liu YY; College of Chemistry and Chemical Engineering, Xinyang Normal University, Xinyang, P. R. China.
  • Zhang JJ; College of Chemistry and Chemical Engineering, Xinyang Normal University, Xinyang, P. R. China.
  • Zhou M; College of Chemistry and Chemical Engineering, Xinyang Normal University, Xinyang, P. R. China.
  • Huang KJ; College of Chemistry and Chemical Engineering, Xinyang Normal University, Xinyang, P. R. China.
  • Chen YH; Xinyang Central Hospital, Xinyang, P. R. China.
  • Ren SW; Xinyang Central Hospital, Xinyang, P. R. China.
Electrophoresis ; 36(19): 2413-8, 2015 Oct.
Article en En | MEDLINE | ID: mdl-26095306
ABSTRACT
A novel aptamer-based CE with chemiluminescence (CL) assay was developed for highly sensitive detection of human immunoglobulin E (IgE). The IgE aptamer was conjugated with gold nanoparticles (AuNPs) to form AuNPs-aptamer that could specifically recognize the IgE to produce an AuNPs-aptamer-IgE complex. The mixture of the AuNPs-aptamer-IgE complex and the unbounded AuNPs-aptamer could be effectively separated by CE and sensitively detected with luminol-H2 O2 CL system. By taking the advantage of the excellent catalytic behavior of AuNPs on luminol-H2 O2 CL system, the ultrasensitive detection of IgE was achieved. The detection limit of IgE is 7.6 fM (S/N = 3) with a linear range from 0.025 to 250 pM. Successful detection of IgE in human serum samples was demonstrated and the recoveries of 94.9-103.2% were obtained. The excellent assay features of the developed approach are its specificity, sensitivity, adaptability, and very small sample consumption. Our design provides a methodology model for determination of rare proteins in biological samples.
Asunto(s)
Palabras clave

Texto completo: 1 Colección: 01-internacional Banco de datos: MEDLINE Asunto principal: Inmunoglobulina E / Electroforesis Capilar / Aptámeros de Nucleótidos / Mediciones Luminiscentes Tipo de estudio: Diagnostic_studies / Prognostic_studies Límite: Humans Idioma: En Revista: Electrophoresis Año: 2015 Tipo del documento: Article

Texto completo: 1 Colección: 01-internacional Banco de datos: MEDLINE Asunto principal: Inmunoglobulina E / Electroforesis Capilar / Aptámeros de Nucleótidos / Mediciones Luminiscentes Tipo de estudio: Diagnostic_studies / Prognostic_studies Límite: Humans Idioma: En Revista: Electrophoresis Año: 2015 Tipo del documento: Article