Analysis of load distribution in tooth-implant supported fixed partial dentures by the use of resilient abutment.
Srp Arh Celok Lek ; 144(3-4): 188-95, 2016.
Article en En | MEDLINE | ID: mdl-27483564
ABSTRACT

INTRODUCTION:

Differences between the tooth and implant response to load can lead to many biological and technical implications in the conditions of occlusal forces.

OBJECTIVE:

The objective of this study was to analyze load distribution in tooth/implant-supported fixed partial dentures with the use of resilient TSA (Titan Shock Absorber, BoneCare GmbH, Augsburg, Germany) abutment and conventional non-resilient abutment using finite element method.

METHODS:

This study presents two basic 3D models. For one model a standard non-resilient abutment is used, and on the implant of the second model a resilient TSA abutment is applied. The virtual model contains drawn contours of tooth, mucous membranes, implant, cortical bones and spongiosa, abutment and suprastructure. The experiment used 500 N of vertical force, applied in three different cases of axial load. Calculations of von Mises equivalent stresses of the tooth root and periodontium, implants and peri-implant tissue were made.

RESULTS:

For the model to which a non-resilient abutment is applied, maximum stress values in all three cases are observed in the cortical part of the bone (maximum stress value of 49.7 MPa). Measurements of stress and deformation in the bone tissue in the model with application of the resilientTSA abutment demonstrated similar distribution; however, these values are many times lower than in the model with non-resilient TSA abutment (maximum stress value of 28.9 MPa).

CONCLUSION:

Application of the resilient TSA abutment results in more equal distribution of stress and deformations in the bone tissue under vertical forces. These values are many times lower than in the model with the non-resilient abutment.
Asunto(s)
Buscar en Google
Colección: 01-internacional Banco de datos: MEDLINE Asunto principal: Fuerza de la Mordida / Implantes Dentales / Pilares Dentales / Soporte de Peso / Prótesis Dental de Soporte Implantado / Dentadura Parcial Fija Límite: Humans Idioma: En Revista: Srp Arh Celok Lek Año: 2016 Tipo del documento: Article
Buscar en Google
Colección: 01-internacional Banco de datos: MEDLINE Asunto principal: Fuerza de la Mordida / Implantes Dentales / Pilares Dentales / Soporte de Peso / Prótesis Dental de Soporte Implantado / Dentadura Parcial Fija Límite: Humans Idioma: En Revista: Srp Arh Celok Lek Año: 2016 Tipo del documento: Article