Air filled, including "air-charged," catheters in urodynamic studies: does the evidence justify their use?
Abrams, Paul; Damaser, Margot S; Niblett, Philip; Rosier, Peter F W M; Toozs-Hobson, Philip; Hosker, Gordon; Kightley, Robert; Gammie, Andrew.
Afiliación
  • Abrams P; Bristol Urological Institute, Southmead Hospital, Bristol, UK.
  • Damaser MS; Department of Biomed Engineering, Cleveland Clinic, Cleveland, Ohio.
  • Niblett P; Research Service, Louis Stokes Cleveland Veterans Administration Medical Center, Cleveland, Ohio.
  • Rosier PFWM; Exeter Scientific Developments, Exeter, UK.
  • Toozs-Hobson P; University Medical Centre Utrecht, Heidelberglaan, Utrecht, The Netherlands.
  • Hosker G; Birmingham Women's NHS Foundation Trust, Birmingham, West Midlands, UK.
  • Kightley R; (Retired) The Warrell Unit, St Marys Hospital, Manchester, UK.
  • Gammie A; Department of Medical Physics and Biomedical Engineering, University College London Hospitals NHS Foundation Trust, London, UK.
Neurourol Urodyn ; 36(5): 1234-1242, 2017 Jun.
Article en En | MEDLINE | ID: mdl-27580083
AIMS: Air filled catheters (AFCs) have been actively marketed for the past few years and in some geographic areas are widely used. However, as the scientific basis for introduction of this technology for pressure measurement in urodynamics was not clear, a study group examined the evidence. METHODS: A search of the peer reviewed literature was carried out. RESULTS: Four papers were identified, of which two were laboratory experiments and two were clinical papers, in female patients, that compared the pressures recorded by AFCs and those recorded using the traditional water filled catheters (WFCs). These data show that there are differences between the pressures measured by the two types of catheters. As yet, the reasons for these differences are not clear. CONCLUSIONS: There should be further systematic laboratory and clinical research before AFCs can be recommended for routine clinical use. We would recommend that a professional worldwide multidisciplinary scientific society, such as the International Continence Society, should work with manufacturers and regulatory bodies to ensure that this urodynamic method is properly scientifically evaluated, in the wider interests of patient safety.
Asunto(s)
Palabras clave

Texto completo: 1 Colección: 01-internacional Banco de datos: MEDLINE Asunto principal: Urodinámica / Aire / Catéteres Límite: Humans Idioma: En Revista: Neurourol Urodyn Año: 2017 Tipo del documento: Article

Texto completo: 1 Colección: 01-internacional Banco de datos: MEDLINE Asunto principal: Urodinámica / Aire / Catéteres Límite: Humans Idioma: En Revista: Neurourol Urodyn Año: 2017 Tipo del documento: Article