Harmonization Study Between LC-MS/MS and Diasorin RIA for Measurement of 25-Hydroxyvitamin D Concentrations in a Large Population Survey.
Berry, Diane J; Dutton, John; Fraser, William D; Järvelin, Marjo-Riitta; Hyppönen, Elina.
Afiliación
  • Berry DJ; Population, Policy and Practice, UCL Institute of Child Health, London, UK.
  • Dutton J; Department of Clinical Biochemistry, Royal Liverpool University Hospital, Liverpool, UK.
  • Fraser WD; Norwich Medical School, University of East Anglia, Norwich Research Park, Norwich, UK.
  • Järvelin MR; Department of Epidemiology and Biostatistics, MRC Health Protection Agency (HPA) Centre for Environment and Health, School of Public Health, Imperial College London, London, UK.
  • Hyppönen E; Unit of Primary Care, Oulu University Hospital, Oulu, Finland.
J Clin Lab Anal ; 31(3)2017 May.
Article en En | MEDLINE | ID: mdl-27595769
ABSTRACT

BACKGROUND:

Population-based research on vitamin D has increased dramatically in recent years. Such studies are typically reliant on assay procedures to measure reliable and comparable levels of 25-hydroxyvitamin D [25(OH)D] concentrations.

METHODS:

Concentrations of 25(OH)D3 and 25(OH)D2 were measured using LC-MS/MS in 5,915 participants (aged 31 years) of Northern Finland Birth Cohort 1966. Blood samples were assayed in batches over a course of 18 months. As anomalies were present in the measurements, 200 samples were reassayed using Diasorin RIA. Agreement between measurements was assessed by Passing-Bablok regression and limits of agreement (LoA). To harmonize LC-MS/MS with Diasorin RIA measurements, formulae were derived from the LoA.

RESULTS:

Concentrations measured by LC-MS/MS were much higher than those measured by Diasorin RIA, with a mean difference of 12.9 ng/ml. Constant variation was evident between batch measurements after log transformation. Statistical formula was applied separately for each batch of LC-MS/MS measurements, enabling us to remove both the constant and proportional bias that was evident prior to the transformation.

CONCLUSION:

Despite the introduction of schemes/programs to improve accuracy of assays to measure 25(OH)D, significant differences can still happen. In these instances, methods to harmonize measurements based on a relatively small number of replicates can be successfully applied to establish confidence and to enable between-study comparisons.
Asunto(s)
Palabras clave

Texto completo: 1 Colección: 01-internacional Banco de datos: MEDLINE Asunto principal: Vitamina D / Radioinmunoensayo / Tamizaje Masivo / Cromatografía Liquida / Espectrometría de Masas en Tándem Tipo de estudio: Etiology_studies / Incidence_studies / Observational_studies / Prognostic_studies / Risk_factors_studies / Screening_studies Límite: Humans País/Región como asunto: Europa Idioma: En Revista: J Clin Lab Anal Asunto de la revista: TECNICAS E PROCEDIMENTOS DE LABORATORIO Año: 2017 Tipo del documento: Article País de afiliación: Reino Unido

Texto completo: 1 Colección: 01-internacional Banco de datos: MEDLINE Asunto principal: Vitamina D / Radioinmunoensayo / Tamizaje Masivo / Cromatografía Liquida / Espectrometría de Masas en Tándem Tipo de estudio: Etiology_studies / Incidence_studies / Observational_studies / Prognostic_studies / Risk_factors_studies / Screening_studies Límite: Humans País/Región como asunto: Europa Idioma: En Revista: J Clin Lab Anal Asunto de la revista: TECNICAS E PROCEDIMENTOS DE LABORATORIO Año: 2017 Tipo del documento: Article País de afiliación: Reino Unido