Your browser doesn't support javascript.
loading
Routine evaluation of left ventricular function using CZT-SPECT, with low injected activities and limited recording times.
Claudin, Marine; Imbert, Laetitia; Djaballah, Wassila; Veran, Nicolas; Poussier, Sylvain; Roch, Véronique; Perrin, Mathieu; Verger, Antoine; Boutley, Henri; Karcher, Gilles; Marie, Pierre-Yves.
Afiliación
  • Claudin M; Department of Nuclear Medicine, CHU-Nancy, Nancy, France.
  • Imbert L; Department of Nuclear Medicine, CHU-Nancy, Nancy, France. l.imbert@chu-nancy.fr.
  • Djaballah W; Nancyclotep Molecular Imaging Platform, Nancy, France. l.imbert@chu-nancy.fr.
  • Veran N; UMR 7039 CRAN and Université de Lorraine, Nancy, France. l.imbert@chu-nancy.fr.
  • Poussier S; Department of Nuclear Medicine, CHU-Nancy, Nancy, France.
  • Roch V; INSERM U947 and Université de Lorraine, Nancy, France.
  • Perrin M; Department of Nuclear Medicine, CHU-Nancy, Nancy, France.
  • Verger A; Nancyclotep Molecular Imaging Platform, Nancy, France.
  • Boutley H; Nancyclotep Molecular Imaging Platform, Nancy, France.
  • Karcher G; INSERM U947 and Université de Lorraine, Nancy, France.
  • Marie PY; Department of Nuclear Medicine, CHU-Nancy, Nancy, France.
J Nucl Cardiol ; 25(1): 249-256, 2018 Feb.
Article en En | MEDLINE | ID: mdl-27677613
ABSTRACT

PURPOSE:

Gamma-cameras, with Cadmium-Zinc-Telluride (CZT) detectors, allow to perform myocardial perfusion imaging (MPI) with limited injected activities and recorded times. This study aimed at determining whether the routine assessment of left ventricular (LV) function with such limited counts protocols compares well with reference values from cardiac MRI.

METHODS:

The study included patients who have undergone cardiac MRI and an MPI routinely planned on a CZT camera with a low-dose protocol (120 MBq of Sestamibi for stress and 360 MBq at rest for 75 kg body weight), while targeting the recording of only 500 myocardial kcounts in order to limit the recording times (<10 minutes for stress, <4 minutes for rest). SPECT images were reconstructed with a method maintaining rather high spatial (8 mm) and temporal (16 frames/cycle) resolutions.

RESULTS:

Seventy-six patients were included, and mean effective dose was 3.5 ± 1.7 mSv for the total MPI protocol. Correlations between CZT-SPECT and MRI were good to excellent for ejection fraction (r 2 = 0.77), end-diastolic (r 2 = 0.88) and end-systolic (r 2 = 0.93) volumes, and the analysis of segmental contractility correlated well between the two techniques (kappa score = 0.72 ± 0.02).

CONCLUSION:

LV function, assessed on a CZT camera with low injected activities and limited recording times, correlates well with the reference assessment from cardiac MRI.
Asunto(s)
Palabras clave

Texto completo: 1 Colección: 01-internacional Banco de datos: MEDLINE Asunto principal: Telurio / Zinc / Cadmio / Imagen por Resonancia Magnética / Tomografía Computarizada de Emisión de Fotón Único / Función Ventricular Izquierda / Imagen de Perfusión Miocárdica Tipo de estudio: Observational_studies Límite: Aged / Female / Humans / Male / Middle aged Idioma: En Revista: J Nucl Cardiol Asunto de la revista: CARDIOLOGIA Año: 2018 Tipo del documento: Article País de afiliación: Francia

Texto completo: 1 Colección: 01-internacional Banco de datos: MEDLINE Asunto principal: Telurio / Zinc / Cadmio / Imagen por Resonancia Magnética / Tomografía Computarizada de Emisión de Fotón Único / Función Ventricular Izquierda / Imagen de Perfusión Miocárdica Tipo de estudio: Observational_studies Límite: Aged / Female / Humans / Male / Middle aged Idioma: En Revista: J Nucl Cardiol Asunto de la revista: CARDIOLOGIA Año: 2018 Tipo del documento: Article País de afiliación: Francia