Capillarys 2 Flex Piercing: Analytical performance assessment according to CLSI protocols for HbA1c quantification.
Doggui, Radhouene; Abdelhafidh Sahli, Chaïma; Aissa, Wassef Lotfi; Hammami, Maroua; Ben Sedrine, Maha; Mahjoub, Rahma; Zouaoui, Khemais; Daboubi, Rim; Siala, Hajer; Messaoud, Taieb; Bibi, Amina.
Afiliación
  • Doggui R; SURVEN Laboratory, National Institute of Nutrition and Food Technology, Tunis, Tunisia.
  • Abdelhafidh Sahli C; Faculty of Sciences of Bizerte, Jarzouna, Tunisia.
  • Aissa WL; Laboratory of Clinical Biology, National Institute of Nutrition and Food Technology, Tunis, Tunisia.
  • Hammami M; Laboratory of Clinical Biology, National Institute of Nutrition and Food Technology, Tunis, Tunisia.
  • Ben Sedrine M; Laboratory of Clinical Biology, National Institute of Nutrition and Food Technology, Tunis, Tunisia.
  • Mahjoub R; Laboratory of Clinical Biology, National Institute of Nutrition and Food Technology, Tunis, Tunisia.
  • Zouaoui K; Laboratory of Clinical Biology, National Institute of Nutrition and Food Technology, Tunis, Tunisia.
  • Daboubi R; Laboratory of Clinical Biology, National Institute of Nutrition and Food Technology, Tunis, Tunisia.
  • Siala H; Laboratory of Biochemistry, Children Hospital, Tunis, Tunisia.
  • Messaoud T; Laboratory of Biochemistry, Children Hospital, Tunis, Tunisia.
  • Bibi A; Laboratory of Biochemistry, Children Hospital, Tunis, Tunisia.
Electrophoresis ; 38(17): 2210-2218, 2017 09.
Article en En | MEDLINE | ID: mdl-28543614
ABSTRACT
HbA1c is used for monitoring diabetic balance. In this paper we report an assessment of the analytical performances of Capillarys 2 Flex Piercing (C2FP) for HbA1c measurement using CE (Capillary Electrophoresis). CLSI (Clinical and Laboratory Standard Institute) protocols are used for the evaluation of apparatus performances precision, linearity, method comparison, trueness and common interferences. HbA1c CVs average in intra-assay was 1.6% between run imprecision CV ranged from 0.1 to 1.8%. The linearity was demonstrated between 4.7 and 15.0%. The comparison study revealed that Bland Altman plot mean difference was equal to -0.03 (CI 95% (-0.05 to -0.0003)) and Passing-Bablok regression intercept was -0.05, CI95%(-0.13 -  -0.05); slope 1.00, CI95%[1.00-1.01]. A strong correlation (r > 0.99) was proved. No significant effects of hemoglobin variants were seen with CE on HbA1c measurement. No problem related to sample-to-sample carry over was noted. No interferences of LA1c and cHb were observed. CE allowed quantification of HbA1c even at low level of total hemoglobin (40 g/L) in contrast to HPLC. Furthermore, this analyzer offered the opportunity of quantifying the HbA2 simultaneously with HbA1c . This evaluation showed that C2FP is a convenient system for the control of diabetes and the detection of hemoglobinopathies.
Asunto(s)
Palabras clave

Texto completo: 1 Colección: 01-internacional Banco de datos: MEDLINE Asunto principal: Hemoglobina Glucada / Electroforesis Capilar Tipo de estudio: Guideline / Prognostic_studies Límite: Humans País/Región como asunto: Africa Idioma: En Revista: Electrophoresis Año: 2017 Tipo del documento: Article País de afiliación: Túnez

Texto completo: 1 Colección: 01-internacional Banco de datos: MEDLINE Asunto principal: Hemoglobina Glucada / Electroforesis Capilar Tipo de estudio: Guideline / Prognostic_studies Límite: Humans País/Región como asunto: Africa Idioma: En Revista: Electrophoresis Año: 2017 Tipo del documento: Article País de afiliación: Túnez