Internal-external malalignment of the femoral component in kinematically aligned total knee arthroplasty increases tibial force imbalance but does not change laxities of the tibiofemoral joint.
Riley, Jeremy; Roth, Joshua D; Howell, Stephen M; Hull, Maury L.
Afiliación
  • Riley J; Biomedical Engineering Graduate Group, University of California, Davis, USA.
  • Roth JD; Biomedical Engineering Graduate Group, University of California, Davis, USA.
  • Howell SM; Department of Biomedical Engineering, University of California, Davis, USA.
  • Hull ML; Department of Biomedical Engineering, University of California, Davis, USA. mlhull@ucdavis.edu.
Knee Surg Sports Traumatol Arthrosc ; 26(6): 1618-1628, 2018 Jun.
Article en En | MEDLINE | ID: mdl-29181558
ABSTRACT

PURPOSE:

The purposes of this study were to quantify the increase in tibial force imbalance (i.e. magnitude of difference between medial and lateral tibial forces) and changes in laxities caused by  2° and 4° of internal-external (I-E) malalignment of the femoral component in kinematically aligned total knee arthroplasty. Because I-E malalignment would introduce the greatest changes to the articular surfaces near 90° of flexion, the hypotheses were that the tibial force imbalance would be significantly increased near 90° flexion and that primarily varus-valgus laxity would be affected near 90° flexion.

METHODS:

Kinematically aligned TKA was performed on ten human cadaveric knee specimens using disposable manual instruments without soft tissue release. One 3D-printed reference femoral component, with unmodified geometry, was aligned to restore the native distal and posterior femoral joint lines. Four 3D-printed femoral components, with modified geometry, introduced I-E malalignments of 2° and 4° from the reference component. Medial and lateral tibial forces were measured from 0° to 120° flexion using a custom tibial force sensor. Bidirectional laxities in four degrees of freedom were measured from 0° to 120° flexion using a custom load application system.

RESULTS:

Tibial force imbalance increased the greatest at 60° flexion where a regression analysis against the degree of I-E malalignment yielded sensitivities (i.e. slopes) of 30 N/° (medial tibial force > lateral tibial force) and 10 N/° (lateral tibial force > medial tibial force) for internal and external malalignments, respectively. Valgus laxity increased significantly with the 4° external component with the greatest increase of 1.5° occurring at 90° flexion (p < 0.0001).

CONCLUSION:

With the tibial component correctly aligned, I-E malalignment of the femoral component caused significant increases in tibial force imbalance. Minimizing I-E malalignment lowers the increase in the tibial force imbalance. By keeping the resection thickness of each posterior femoral condyle to within ± 0.5 mm of the thickness of the respective posterior region of the femoral component, the increase in imbalance can be effectively limited to 38 N. Generally laxities were unaffected within the ± 4º range tested indicating that instability is not a clinical concern and that manual testing of laxities is not useful to detect I-E malalignment.
Asunto(s)
Palabras clave

Texto completo: 1 Colección: 01-internacional Banco de datos: MEDLINE Asunto principal: Desviación Ósea / Artroplastia de Reemplazo de Rodilla / Inestabilidad de la Articulación / Articulación de la Rodilla / Prótesis de la Rodilla Tipo de estudio: Guideline Límite: Aged / Aged80 / Female / Humans / Male / Middle aged Idioma: En Revista: Knee Surg Sports Traumatol Arthrosc Asunto de la revista: MEDICINA ESPORTIVA / TRAUMATOLOGIA Año: 2018 Tipo del documento: Article País de afiliación: Estados Unidos

Texto completo: 1 Colección: 01-internacional Banco de datos: MEDLINE Asunto principal: Desviación Ósea / Artroplastia de Reemplazo de Rodilla / Inestabilidad de la Articulación / Articulación de la Rodilla / Prótesis de la Rodilla Tipo de estudio: Guideline Límite: Aged / Aged80 / Female / Humans / Male / Middle aged Idioma: En Revista: Knee Surg Sports Traumatol Arthrosc Asunto de la revista: MEDICINA ESPORTIVA / TRAUMATOLOGIA Año: 2018 Tipo del documento: Article País de afiliación: Estados Unidos