Counting-loss correction for X-ray spectroscopy using unit impulse pulse shaping.
Hong, Xu; Zhou, Jianbin; Ni, Shijun; Ma, Yingjie; Yao, Jianfeng; Zhou, Wei; Liu, Yi; Wang, Min.
Afiliación
  • Hong X; College of Nuclear Technology and Automation Engineering, Chengdu University of Technology, No.1 the Third Section East, Er Xianqiao Road, Chengdu, Sichuan 610059, People's Republic of China.
  • Zhou J; College of Nuclear Technology and Automation Engineering, Chengdu University of Technology, No.1 the Third Section East, Er Xianqiao Road, Chengdu, Sichuan 610059, People's Republic of China.
  • Ni S; College of Nuclear Technology and Automation Engineering, Chengdu University of Technology, No.1 the Third Section East, Er Xianqiao Road, Chengdu, Sichuan 610059, People's Republic of China.
  • Ma Y; College of Nuclear Technology and Automation Engineering, Chengdu University of Technology, No.1 the Third Section East, Er Xianqiao Road, Chengdu, Sichuan 610059, People's Republic of China.
  • Yao J; College of Nuclear Technology and Automation Engineering, Chengdu University of Technology, No.1 the Third Section East, Er Xianqiao Road, Chengdu, Sichuan 610059, People's Republic of China.
  • Zhou W; College of Nuclear Technology and Automation Engineering, Chengdu University of Technology, No.1 the Third Section East, Er Xianqiao Road, Chengdu, Sichuan 610059, People's Republic of China.
  • Liu Y; College of Nuclear Technology and Automation Engineering, Chengdu University of Technology, No.1 the Third Section East, Er Xianqiao Road, Chengdu, Sichuan 610059, People's Republic of China.
  • Wang M; College of Nuclear Technology and Automation Engineering, Chengdu University of Technology, No.1 the Third Section East, Er Xianqiao Road, Chengdu, Sichuan 610059, People's Republic of China.
J Synchrotron Radiat ; 25(Pt 2): 505-513, 2018 Mar 01.
Article en En | MEDLINE | ID: mdl-29488930
ABSTRACT
High-precision measurement of X-ray spectra is affected by the statistical fluctuation of the X-ray beam under low-counting-rate conditions. It is also limited by counting loss resulting from the dead-time of the system and pile-up pulse effects, especially in a high-counting-rate environment. In this paper a detection system based on a FAST-SDD detector and a new kind of unit impulse pulse-shaping method is presented, for counting-loss correction in X-ray spectroscopy. The unit impulse pulse-shaping method is evolved by inverse deviation of the pulse from a reset-type preamplifier and a C-R shaper. It is applied to obtain the true incoming rate of the system based on a general fast-slow channel processing model. The pulses in the fast channel are shaped to unit impulse pulse shape which possesses small width and no undershoot. The counting rate in the fast channel is corrected by evaluating the dead-time of the fast channel before it is used to correct the counting loss in the slow channel.