Interchangeability of Procalcitonin Measurements Using the Point of Care Testing i-CHROMATM Reader and the Automated Liaison XL.
Clin Lab ; 64(6): 1097-1100, 2018 Jun 01.
Article en En | MEDLINE | ID: mdl-29945312
ABSTRACT

BACKGROUND:

Our aim was to verify if procalcitonin (PCT) measurements using the new point-of-care testing i-CHROMATM are interchangeable with those of Liaison XL.

METHODS:

One hundred seventeen serum samples were processed sequentially on a Liaison XL and i-CHROMATM. Statistical analysis was done using the Passing-Bablok regression, Bland-Altman test, and Cohen's Kappa statistic.

RESULTS:

Proportional and constant differences were observed between i-CHROMATM and Liaison XL. The 95% CI of the mean bias% was very large, exceeding the maximum allowable TE% and the clinical reference change value. However, the concordance between methods at the clinical relevant cutoffs was strong, with the exception of the 0.25 ng/mL cutoff which was moderate.

CONCLUSIONS:

Our data suggest that i-CHROMATM is not interchangeable with Liaison XL. However, while the strong concordance at the clinical relevant cutoffs allows us to consider i-CHROMATM a suitable option to Liaison XL to support clinicians' decision-making; nevertheless, the moderate agreement at the 0.25 ng/mL cutoff recommends caution in interpreting the data around this cutoff.
Asunto(s)

Texto completo: 1 Colección: 01-internacional Banco de datos: MEDLINE Asunto principal: Automatización de Laboratorios / Pruebas en el Punto de Atención / Polipéptido alfa Relacionado con Calcitonina Tipo de estudio: Diagnostic_studies / Prognostic_studies Límite: Humans Idioma: En Revista: Clin Lab Asunto de la revista: TECNICAS E PROCEDIMENTOS DE LABORATORIO Año: 2018 Tipo del documento: Article

Texto completo: 1 Colección: 01-internacional Banco de datos: MEDLINE Asunto principal: Automatización de Laboratorios / Pruebas en el Punto de Atención / Polipéptido alfa Relacionado con Calcitonina Tipo de estudio: Diagnostic_studies / Prognostic_studies Límite: Humans Idioma: En Revista: Clin Lab Asunto de la revista: TECNICAS E PROCEDIMENTOS DE LABORATORIO Año: 2018 Tipo del documento: Article