Your browser doesn't support javascript.
loading
Minimizing Endothelial Cell Loss Caused by Orientation Stamps on Preloaded Descemet Membrane Endothelial Keratoplasty Grafts.
Newman, Lara R; Tran, Khoa D; Odell, Kelly; Dye, Philip K; Galloway, Joshua; Sales, Christopher S; Straiko, Michael D; Terry, Mark A.
Afiliación
  • Newman LR; Central Pennsylvania Eye Institute, Hershey, PA.
  • Tran KD; Lions VisionGift, Portland, OR.
  • Odell K; Lions VisionGift, Portland, OR.
  • Dye PK; Lions VisionGift, Portland, OR.
  • Galloway J; Lions VisionGift, Portland, OR.
  • Sales CS; Weill Cornell Medical College, New York, NY.
  • Straiko MD; Lions VisionGift, Portland, OR.
  • Terry MA; Cornea Service, Devers Eye Institute, Portland, OR.
Cornea ; 38(2): 233-237, 2019 Feb.
Article en En | MEDLINE | ID: mdl-30418274
ABSTRACT

PURPOSE:

To quantify endothelial cell loss (ECL) caused by orientation stamps on prestripped and preloaded Descemet membrane endothelial keratoplasty (DMEK) grafts, and to examine a method for reducing ECL using a smaller stamp.

METHODS:

Ten prestripped and 10 preloaded DMEK grafts were prepared with S-stamps. Ten additional preloaded DMEK grafts were prepared with both an S-stamp and a smaller F-stamp in different paracentral areas of the graft. The footprint of each stamp was measured using ink on cardstock. DMEK grafts were stored in viewing chambers filled with 20 mL of Optisol-GS for 3 days at 4°C. ECL was quantified using Calcein-AM staining and FIJI Weka Segmentation.

RESULTS:

S-stamps on prestripped DMEK grafts contributed an average ECL of 1.1% ± 0.5% (range 0.6%-2.2%) toward total graft damage, whereas S-stamps on preloaded DMEK grafts contributed approximately twice that amount (average ECL 2.0% ± 0.7%, range 1.3%-3.1%, P = 0.004). Overall ECL for prestripped grafts (average 7.1% ± 3.3%, range 3.3%-13.7%) and preloaded grafts (average 11.3% ± 4.2%, range 6.9%-19.4%) was similar to previous reports. The footprint of the S-stamp was approximately 45% larger than that of the F-stamp. In 10 preloaded grafts marked with both stamps, the S-stamp caused an average ECL of 1.9% ± 0.6% (range 1.2%-3.2%), whereas the smaller F-stamp caused an average ECL of 1.0% ± 0.2% (range 0.8%-1.4%, P = 0.0002).

CONCLUSIONS:

Loss of endothelial cells associated with graft-stamping was greater in preloaded tissue than in prestripped tissue and was less with a smaller F-stamp than with a larger S-stamp. Using a smaller stamp could help minimize ECL in prestripped and preloaded DMEK grafts.
Asunto(s)

Texto completo: 1 Colección: 01-internacional Banco de datos: MEDLINE Asunto principal: Recolección de Tejidos y Órganos / Bancos de Ojos / Pérdida de Celulas Endoteliales de la Córnea / Queratoplastia Endotelial de la Lámina Limitante Posterior Límite: Aged / Female / Humans / Male / Middle aged Idioma: En Revista: Cornea Año: 2019 Tipo del documento: Article País de afiliación: Panamá

Texto completo: 1 Colección: 01-internacional Banco de datos: MEDLINE Asunto principal: Recolección de Tejidos y Órganos / Bancos de Ojos / Pérdida de Celulas Endoteliales de la Córnea / Queratoplastia Endotelial de la Lámina Limitante Posterior Límite: Aged / Female / Humans / Male / Middle aged Idioma: En Revista: Cornea Año: 2019 Tipo del documento: Article País de afiliación: Panamá