Your browser doesn't support javascript.
loading
Accuracy and Precision of the SynchroMed II Pump.
Farid, Reza; Binz, Kirstin; Emerson, Jane Anne; Murdock, Fred.
Afiliación
  • Farid R; Department of Physical Medicine and Rehabilitation, University of Missouri, Columbia, MO, USA.
  • Binz K; Department of Physical Medicine and Rehabilitation, University of Missouri, Columbia, MO, USA.
  • Emerson JA; Department of Physical Medicine and Rehabilitation, University of Missouri, Columbia, MO, USA.
  • Murdock F; Department of Physical Medicine and Rehabilitation, University of Missouri, Columbia, MO, USA.
Neuromodulation ; 22(7): 805-810, 2019 Oct.
Article en En | MEDLINE | ID: mdl-30889307
ABSTRACT

OBJECTIVE:

The objective of this study was to evaluate the accuracy and precision of SynchroMed II programmable infusion pumps for delivery of intrathecal baclofen to treat severe spasticity.

DESIGN:

Retrospective chart review study.

SETTING:

Urban academic rehabilitation clinic. PATIENTS (OR PARTICIPANTS) Patients with spasticity treated with intrathecal baclofen. METHODS OR INTERVENTIONS (OR ASSESSMENT OF RISK FACTORS) A retrospective chart review of 149 patients inclusive of 755 individual patient encounters more than a 2.5-year span for baclofen pump refills was performed. MAIN OUTCOME MEASUREMENTS Charts were reviewed to obtain data on the concentration and dose of intrathecal baclofen, aspirated and residual volumes, age of the pump, and reservoir volume of the pump. Flow rate (mL/day) was calculated for each chronological visit. Flow rates from clinical data were used to calculate the accuracy and precision based on the manufacturer's specifications for both the 20 mL and 40 mL pumps.

RESULTS:

The infused volume accuracy and precision of both the 20 mL pumps and 40 mL pumps remained within manufacturer's specifications while delivering drug at a rate slightly lower than predicted. There were no cases of pump over-infusion. There is a statistically significant decrease in flow rate accuracy over the lifespan of the pump.

CONCLUSIONS:

SynchroMed II pumps infuse at slightly less than their predicted rate while remaining within the manufacturer's specifications. No pump over-infusions were noted. Pump flow rate declines over time. Further study is warranted to assess the impact of this flow rate decrease on patient treatment and safety.
Asunto(s)
Palabras clave

Texto completo: 1 Colección: 01-internacional Banco de datos: MEDLINE Asunto principal: Baclofeno / Bombas de Infusión Implantables / Relajantes Musculares Centrales / Espasticidad Muscular Tipo de estudio: Diagnostic_studies / Observational_studies / Prognostic_studies / Risk_factors_studies Límite: Humans Idioma: En Revista: Neuromodulation Año: 2019 Tipo del documento: Article País de afiliación: Estados Unidos

Texto completo: 1 Colección: 01-internacional Banco de datos: MEDLINE Asunto principal: Baclofeno / Bombas de Infusión Implantables / Relajantes Musculares Centrales / Espasticidad Muscular Tipo de estudio: Diagnostic_studies / Observational_studies / Prognostic_studies / Risk_factors_studies Límite: Humans Idioma: En Revista: Neuromodulation Año: 2019 Tipo del documento: Article País de afiliación: Estados Unidos