Your browser doesn't support javascript.
loading
Transmural unipolar electrogram change occurs within 7 s at the left atrial posterior wall during pulmonary vein isolation.
Tomlinson, David R; Myles, Madison; Stevens, Kara N; Streeter, Adam J.
Afiliación
  • Tomlinson DR; South West Cardiothoracic Centre, University Hospitals Plymouth NHS Trust, Plymouth, UK.
  • Myles M; Dorset Heart Centre, Royal Bournemouth and Christchurch Hospitals NHS Foundation, Bournemouth, UK.
  • Stevens KN; Department of Medical Statistics, Plymouth University Peninsula Schools of Medicine and Dentistry, Plymouth, UK.
  • Streeter AJ; Department of Medical Statistics, Plymouth University Peninsula Schools of Medicine and Dentistry, Plymouth, UK.
Pacing Clin Electrophysiol ; 42(7): 922-929, 2019 07.
Article en En | MEDLINE | ID: mdl-31127636
ABSTRACT

BACKGROUND:

To assess occurrence of a histologically validated measure of transmural (TM) atrial ablation-pure R unipolar electrogram (UE) morphology change-at first-ablated left atrial posterior wall (LAPW) sites during contact force (CF)-guided pulmonary vein isolation (PVI).

METHODS:

Objectively annotated VISITAG™ Module and CARTOREPLAY™ (Biosense Webster Inc., Diamond Bar, CA, USA) UE morphology data were retrospectively analyzed in 23 consecutive patients undergoing PVI under general anesthesia.

RESULTS:

PVI without spontaneous/dormant recovery was achieved in all, employing 16.3 (3.2) min of radiofrequency (RF; 30 W) energy. All first-ablated LAPW sites demonstrated RS UE morphology preablation, with RF-induced pure R UE morphology change in 98%. Time to pure R UE morphology was significantly shorter at left-sided LAPW sites (4.9 [2.1] vs 6.7 [2.5] s; P = .02), with significantly greater impedance drop (median 13.5 vs 9.9 Ω; P = .003). Importantly, neither first-site RF duration (14.9 vs 15.0 s) nor maximum ablation catheter tip distance moved (during RF) was significantly different, yet the mean CF was significantly higher at right-sided sites (16.5 vs 11.2 g; P = .002). Concurrent impedance and objectively annotated bipolar electrogram (BE) data demonstrated ∼6-8 Ω impedance drop and ∼30% BE decrease at the time of first pure R UE morphology change.

CONCLUSIONS:

Using objective ablation site annotation, UE morphology evidence of TM RF effect was demonstrated far sooner than considered biologically possible according to the "conventional" 20-40 s RF per-site approach, with significantly greater ablative effect evident at left-sided sites. This novel methodology represents a scientifically more rigorous foundation toward future research into the biological effects of RF ablation in vivo.
Asunto(s)
Palabras clave

Texto completo: 1 Colección: 01-internacional Banco de datos: MEDLINE Asunto principal: Venas Pulmonares / Fibrilación Atrial / Ablación por Catéter / Técnicas Electrofisiológicas Cardíacas Tipo de estudio: Observational_studies Límite: Female / Humans / Male / Middle aged Idioma: En Revista: Pacing Clin Electrophysiol Año: 2019 Tipo del documento: Article País de afiliación: Reino Unido

Texto completo: 1 Colección: 01-internacional Banco de datos: MEDLINE Asunto principal: Venas Pulmonares / Fibrilación Atrial / Ablación por Catéter / Técnicas Electrofisiológicas Cardíacas Tipo de estudio: Observational_studies Límite: Female / Humans / Male / Middle aged Idioma: En Revista: Pacing Clin Electrophysiol Año: 2019 Tipo del documento: Article País de afiliación: Reino Unido