Improve depth of field of optical coherence tomography using finite energy Airy beam.
Opt Lett ; 44(12): 3158-3161, 2019 Jun 15.
Article en En | MEDLINE | ID: mdl-31199405
We report a technique to break the depth of field (DOF) limit in optical coherence tomography (OCT) using a finite energy Airy beam. The Airy beam is generated using a phase mask in a Fourier transform schematic and provides the DOF improvement due to its low diffraction. We compare Airy beam OCT with conventional Gaussian beam OCT using lateral resolution and sensitivity. Experimental data from the polystyrene beads in water as well as lemon tissue confirm the extension of DOF up to 10 mm in Airy beam OCT, while the DOF of Gaussian beam OCT is less than 3.0 mm. We also demonstrate that a modified Airy beam can be effectively used in OCT by adjusting the truncating factor of the Airy beam via changing the pattern scale of the phase mask. This result provides a selection method for the use of a finite energy Airy beam in OCT.

Texto completo: 1 Colección: 01-internacional Banco de datos: MEDLINE Idioma: En Revista: Opt Lett Año: 2019 Tipo del documento: Article

Texto completo: 1 Colección: 01-internacional Banco de datos: MEDLINE Idioma: En Revista: Opt Lett Año: 2019 Tipo del documento: Article