Your browser doesn't support javascript.
loading
2D shear wave elastography: measurement acquisition and reliability criteria in noninvasive assessment of liver fibrosis.
Chung, Maggie; Baird, Grayson L; Weiss, Krista E; Beland, Michael D.
Afiliación
  • Chung M; Department of Radiology and Biomedical Imaging, University of California, San Francisco, 505 Parnassus Ave., Room M391, Box 0628, San Francisco, CA, 94143, USA. Mchung2017@gmail.com.
  • Baird GL; Lifespan Biostatistics Core, Rhode Island Hospital, 593 Eddy St., Providence, RI, 02903, USA.
  • Weiss KE; Department of Diagnostic Imaging, Warren Alpert Medical School of Brown University, Rhode Island Hospital, 593 Eddy St., Providence, RI, 02903, USA.
  • Beland MD; Department of Diagnostic Imaging, Warren Alpert Medical School of Brown University, Rhode Island Hospital, 593 Eddy St., Providence, RI, 02903, USA.
Abdom Radiol (NY) ; 44(10): 3285-3294, 2019 10.
Article en En | MEDLINE | ID: mdl-31435762
ABSTRACT

PURPOSE:

The objective was to evaluate the accuracy of 2D shear wave elastography (SWE) in predicting stages of liver fibrosis using five individual versus grouped measurements and different reliability criteria. MATERIALS AND

METHODS:

This is a prospective study of 109 patients who underwent hepatic 2D SWE (Canon Aplio 500) prior to liver biopsy for varied indications. Liver fibrosis was staged using the METAVIR scoring system (F = 0-4). Propagation mapping was used to guide ten SWE measurements from the liver parenchyma five individual measurements and five grouped measurements. IQR/median, SD/median, and SD/mean were examined as quality criteria for patient inclusion at various thresholds (IQR/median ≤ 0.15, 0.2, 0.3, 0.4, 0.5; SD/median ≤ 0.15, 0.2, 0.3; SD/mean ≤ 0.2, 0.3, 0.5). Threshold for clinically significant fibrosis (F ≥ 2) was determined with receiver operating characteristic (ROC) analysis.

RESULTS:

There was high agreement between individual and grouped measurements without statistically significant differences (intraclass correlation coefficient = 0.82; p = 0.26-0.96). When no quality criterion was used (n = 103), the optimal threshold was 11.3 kPa [AUROC 0.78, 95% CI (0.69, 0.88)] with sensitivity and specificity of 80% and 66%, respectively. All quality criteria were associated with equal or higher AUROC ranging from 0.78 to 0.87. IQR/median ≤ 0.5 (n = 88) achieved the highest sensitivity of 85% and only excluded a small subset of patients. The AUROC and specificity were 0.83 [95% CI (0.74, 0.92)] and 72%, respectively.

SIGNIFICANCE:

Quality criterion IQR/median ≤ 0.5 increases sensitivity and specificity in prediction of clinically significant liver fibrosis while excluding only a small subset of patients. Grouped measurements are comparable to individual measurements and may help increase procedural efficiency.
Asunto(s)
Palabras clave

Texto completo: 1 Colección: 01-internacional Banco de datos: MEDLINE Asunto principal: Diagnóstico por Imagen de Elasticidad / Cirrosis Hepática Tipo de estudio: Diagnostic_studies / Observational_studies / Prognostic_studies / Risk_factors_studies Límite: Adult / Aged / Aged80 / Female / Humans / Male / Middle aged Idioma: En Revista: Abdom Radiol (NY) Año: 2019 Tipo del documento: Article País de afiliación: Estados Unidos

Texto completo: 1 Colección: 01-internacional Banco de datos: MEDLINE Asunto principal: Diagnóstico por Imagen de Elasticidad / Cirrosis Hepática Tipo de estudio: Diagnostic_studies / Observational_studies / Prognostic_studies / Risk_factors_studies Límite: Adult / Aged / Aged80 / Female / Humans / Male / Middle aged Idioma: En Revista: Abdom Radiol (NY) Año: 2019 Tipo del documento: Article País de afiliación: Estados Unidos