Novel implant-free loop Tenodesis vs. simple Tenotomy of the long biceps tendon - a biomechanical investigation.
Kerschbaum, Maximilian; Voss, Andreas; Pfeifer, Christian; Weiss, Isabella; Mayr, Agnes; Alt, Volker; Greiner, Stefan; Grechenig, Stephan.
Afiliación
  • Kerschbaum M; Clinic of Trauma Surgery, University Medical Center Regensburg, Franz-Josef-Strauss-Allee 11, 93053, Regensburg, Germany. Maximilian.kerschbaum@ukr.de.
  • Voss A; Clinic of Trauma Surgery, University Medical Center Regensburg, Franz-Josef-Strauss-Allee 11, 93053, Regensburg, Germany.
  • Pfeifer C; Sporthopaedicum Regensburg, Hildegard von Bingen Strasse 1, 93053, Regensburg, Germany.
  • Weiss I; Clinic of Trauma Surgery, University Medical Center Regensburg, Franz-Josef-Strauss-Allee 11, 93053, Regensburg, Germany.
  • Mayr A; Clinic of Trauma Surgery, University Medical Center Regensburg, Franz-Josef-Strauss-Allee 11, 93053, Regensburg, Germany.
  • Alt V; Clinic of Trauma Surgery, University Medical Center Regensburg, Franz-Josef-Strauss-Allee 11, 93053, Regensburg, Germany.
  • Greiner S; Clinic of Trauma Surgery, University Medical Center Regensburg, Franz-Josef-Strauss-Allee 11, 93053, Regensburg, Germany.
  • Grechenig S; Sporthopaedicum Regensburg, Hildegard von Bingen Strasse 1, 93053, Regensburg, Germany.
BMC Musculoskelet Disord ; 20(1): 522, 2019 Nov 09.
Article en En | MEDLINE | ID: mdl-31706308
ABSTRACT

BACKGROUND:

Simple tenotomy and anchor tenodesis are commonly used in treatment of long biceps tendon (LHB) pathologies. The tenotomy can result in biceps distalization or cosmetic deformities. A novel loop Tenodesis Technique (LTD) could prevent a distalization of the biceps muscle without the risk of implant associated complications. The purpose of this study was to investigate the biomechanical aspects of the novel LTD compared to a standard tenotomy of the LHB. It has been hypothesized that the novel technique will show biomechanical superiority in terms of resistance and distalization.

METHODS:

Seven paired adult human cadaveric shoulder joints were assigned to one of the two study groups Loop tenodesis (LTD); simple tenotomy (STT). In both groups load-to-failure testing was performed. The load-displacement curve was used to determine the maximum load (N), the degree of distalization of the LHB (mm) and the stiffness (N/mm). Additionally, the mode of failure was registered.

RESULTS:

The LTD group achieved a significantly higher ultimate load to failure (LTD 50.5 ± 12.5 N vs. STT 6.6 ± 3.9 N; p = 0.001). Significantly less distalization of the tendon could be detected for the LTD group (LTD 8 ± 2.3 mm vs. STT 22.4 ± 2.4 mm; p = 0.001). Stiffness was 7.4 ± 3.9 N/mm for the LTD group and 0.23 ± 0.16 N/mm for the STT group (p = 0.001). In all specimens of the LTD group a tendon rupture was found as mode of failure, while the STT group failed because of pulling out the LHB through the bicipital groove.

CONCLUSION:

The novel loop Tenodesis Technique shows biomechanically higher stability as well as less distalization compared to a simple tenotomy of the long biceps tendon.
Asunto(s)
Palabras clave

Texto completo: 1 Colección: 01-internacional Banco de datos: MEDLINE Asunto principal: Artroscopía / Músculo Esquelético / Tendinopatía / Tenodesis / Tenotomía Límite: Adult / Humans Idioma: En Revista: BMC Musculoskelet Disord Asunto de la revista: FISIOLOGIA / ORTOPEDIA Año: 2019 Tipo del documento: Article País de afiliación: Alemania

Texto completo: 1 Colección: 01-internacional Banco de datos: MEDLINE Asunto principal: Artroscopía / Músculo Esquelético / Tendinopatía / Tenodesis / Tenotomía Límite: Adult / Humans Idioma: En Revista: BMC Musculoskelet Disord Asunto de la revista: FISIOLOGIA / ORTOPEDIA Año: 2019 Tipo del documento: Article País de afiliación: Alemania