A real-time system for selectively sensing and pacing the His-bundle during sinus rhythm and ventricular fibrillation.
Shah, Ankur R; Khan, Muhammad S; Hirahara, Annie M; Lange, Matthias; Ranjan, Ravi; Dosdall, Derek J.
Afiliación
  • Shah AR; Department of Biomedical Engineering, The University of Utah, Salt Lake City, UT, 84112, USA.
  • Khan MS; Nora Eccles Harrison Cardiovascular Research and Training Institute, The University of Utah, Salt Lake City, UT, 84112, USA.
  • Hirahara AM; Nora Eccles Harrison Cardiovascular Research and Training Institute, The University of Utah, Salt Lake City, UT, 84112, USA.
  • Lange M; Department of Biomedical Engineering, The University of Utah, Salt Lake City, UT, 84112, USA.
  • Ranjan R; Nora Eccles Harrison Cardiovascular Research and Training Institute, The University of Utah, Salt Lake City, UT, 84112, USA.
  • Dosdall DJ; Nora Eccles Harrison Cardiovascular Research and Training Institute, The University of Utah, Salt Lake City, UT, 84112, USA.
Biomed Eng Online ; 19(1): 19, 2020 Apr 10.
Article en En | MEDLINE | ID: mdl-32276597
ABSTRACT

BACKGROUND:

The His-Purkinje (HP) system provides a pathway for the time-synchronous contraction of the heart. His bundle (HB) of the HP system is gaining relevance as a pacing site for treating non-reversible bradyarrhythmia despite limited availability of tools to identify the HB. In this paper, we describe a real-time stimulation and recording system (rt-SRS) to investigate using multi-electrode techniques to identify and selectively pace the HB. The rt-SRS can not only be used in sinus rhythm, but also during ventricular fibrillation (VF). The rt-SRS will also help investigate the so far unknown causal effects of selectively pacing the HB during VF.

METHODS:

The rt-SRS consists of preamplifiers, data acquisition cards interfaced with a real-time controller, a current source, and current routing switches on a remote computer, which may be interrupted to stimulate using a host machine. The remote computer hosts a series of algorithms designed to aid in identifying electrodes directly over the HB, to accurately detect activation rates without over-picking, and to deliver stimulation pulses. The performance of the rt-SRS was demonstrated in seven isolated, perfused rabbit hearts.

RESULTS:

The rt-SRS can visualize up to 96 channels of raw data, and spatial derivative data at 6.25-kHz sampling rate with an input-referred noise of 100 µV. The rt-SRS can send up to ± 150 V of stimuli pulses to any of the 96 channels. In the rabbit experiments, HB activations were detected in 18 ± 6.8% of the 64 electrodes used during VF.

CONCLUSIONS:

The rt-SRS is capable of measuring and responding to cardiac electrophysiological phenomena in real-time with precisely timed and placed electrical stimuli. This rt-SRS was shown to be an effective research tool by successfully detecting and quantifying HB activations and delivering stimulation pulses to selected electrodes in real-time.
Asunto(s)
Palabras clave

Texto completo: 1 Colección: 01-internacional Banco de datos: MEDLINE Asunto principal: Fibrilación Ventricular / Fascículo Atrioventricular / Electrofisiología Límite: Animals Idioma: En Revista: Biomed Eng Online Asunto de la revista: ENGENHARIA BIOMEDICA Año: 2020 Tipo del documento: Article País de afiliación: Estados Unidos

Texto completo: 1 Colección: 01-internacional Banco de datos: MEDLINE Asunto principal: Fibrilación Ventricular / Fascículo Atrioventricular / Electrofisiología Límite: Animals Idioma: En Revista: Biomed Eng Online Asunto de la revista: ENGENHARIA BIOMEDICA Año: 2020 Tipo del documento: Article País de afiliación: Estados Unidos