Effect of patient-specific instrument on lowering threshold for junior physicians to perform total hip arthroplasty on developmental dysplasia of the hip patients.
Xiao, Han; Wang, Chenggong; Zhong, Da; Lei, Pengfei; Hu, Yihe; Su, Shilong.
Afiliación
  • Xiao H; Department of Sports Medicine, Xiangya Hospital, Central South University, Changsha, Hunan, China.
  • Wang C; Digital Orthopaedic Laboratory, Xiangya Hospital, Central South University, Changsha, Hunan, China.
  • Zhong D; Digital Orthopaedic Laboratory, Xiangya Hospital, Central South University, Changsha, Hunan, China.
  • Lei P; Department of Orthopaedics, Xiangya Hospital, Central South University, Changsha, Hunan, China.
  • Hu Y; Digital Orthopaedic Laboratory, Xiangya Hospital, Central South University, Changsha, Hunan, China. Dr_zhongda@126.com.
  • Su S; Department of Orthopaedics, Xiangya Hospital, Central South University, Changsha, Hunan, China. Dr_zhongda@126.com.
Int Orthop ; 44(7): 1281-1286, 2020 07.
Article en En | MEDLINE | ID: mdl-32405884
ABSTRACT

PURPOSE:

To create a patient-specific instrument (PSI) in lowering the surgical experience requirement for junior physicians to perform total hip arthroplasty (THA) on developmental dysplasia of the hip (DDH) patients.

METHODS:

Combined with rapid prototyping technology, we created a PSI and established DDH hip model in vitro. We enrolled 48 junior physicians and randomly assigned them into two groups. After creation of the PSI, they performed simulated THA surgery on a full-scale hip model with or without PSI on DDH models. The planned prothesis orientation, post-operative prothesis orientation, and surgery time were recorded.

RESULTS:

The final cup inclination was 42.0 ± 0.8° in PSI group and 37.8 ± 2.0° in control group, while final cup anteversion was 16.0 ± 0.7° in PSI group and 24.7 ± 3.5° in control group. The △inclination in PSI group was smaller than that in control group (4.2 ± 0.5° vs 9.5 ± 1.4°, P < 0.01), so does △inclination (2.9 ± 0.4° in PSI group vs 15.2 ± 2.5° in control group, P < 0.01). The outlier percent was 8.3% in PSI group and 70.8% in control group (P < 0.01). At the same time, the PSI group did not prolong the operation time (P = 0.551).

CONCLUSION:

The PSI can greatly increase the accuracy of placing the cup orientation and lower the threshold for junior physicians to perform THA on DDH patients. It could be a training tool for them to increase their THA surgical skills.
Asunto(s)
Palabras clave

Texto completo: 1 Colección: 01-internacional Banco de datos: MEDLINE Asunto principal: Artroplastia de Reemplazo de Cadera / Displasia del Desarrollo de la Cadera / Prótesis de Cadera Tipo de estudio: Clinical_trials Límite: Humans Idioma: En Revista: Int Orthop Año: 2020 Tipo del documento: Article País de afiliación: China

Texto completo: 1 Colección: 01-internacional Banco de datos: MEDLINE Asunto principal: Artroplastia de Reemplazo de Cadera / Displasia del Desarrollo de la Cadera / Prótesis de Cadera Tipo de estudio: Clinical_trials Límite: Humans Idioma: En Revista: Int Orthop Año: 2020 Tipo del documento: Article País de afiliación: China