TURBINE-MRE: A 3D hybrid radial-Cartesian EPI acquisition for MR elastography.
Sui, Yi; Arani, Arvin; Trzasko, Joshua D; Murphy, Matthew C; Rossman, Phillip J; Glaser, Kevin J; McGee, Kiaran P; Manduca, Armando; Ehman, Richard L; Araoz, Philip A; Huston, John.
Afiliación
  • Sui Y; Department of Radiology, Mayo Clinic, Rochester, Minnesota, USA.
  • Arani A; Department of Radiology, Mayo Clinic, Rochester, Minnesota, USA.
  • Trzasko JD; Department of Radiology, Mayo Clinic, Rochester, Minnesota, USA.
  • Murphy MC; Department of Radiology, Mayo Clinic, Rochester, Minnesota, USA.
  • Rossman PJ; Department of Radiology, Mayo Clinic, Rochester, Minnesota, USA.
  • Glaser KJ; Department of Radiology, Mayo Clinic, Rochester, Minnesota, USA.
  • McGee KP; Department of Radiology, Mayo Clinic, Rochester, Minnesota, USA.
  • Manduca A; Department of Physiology & Biomedical Engineering, Mayo Clinic, Rochester, Minnesota, USA.
  • Ehman RL; Department of Radiology, Mayo Clinic, Rochester, Minnesota, USA.
  • Araoz PA; Department of Radiology, Mayo Clinic, Rochester, Minnesota, USA.
  • Huston J; Department of Radiology, Mayo Clinic, Rochester, Minnesota, USA.
Magn Reson Med ; 85(2): 945-952, 2021 02.
Article en En | MEDLINE | ID: mdl-32738084
PURPOSE: To develop a novel magnetic resonance elastography (MRE) acquisition using a hybrid radial EPI readout scheme (TURBINE), and to demonstrate its feasibility to obtain wave images and stiffness maps in a phantom and in vivo brain. METHOD: The proposed 3D TURBINE-MRE is based on a spoiled gradient-echo MRE sequence with the EPI readout radially rotating about the phase-encoding axis to sample a full 3D k-space. A polyvinyl chloride phantom and 6 volunteers were scanned on a compact 3T GE scanner with a 32-channel head coil at 80 Hz and 60 Hz external vibration, respectively. For comparison, a standard 2D, multislice, spin-echo (SE) EPI-MRE acquisition was also performed with the same motion encoding and resolution. The TURBINE-MRE images were off-line reconstructed with iterative SENSE algorithm. The regional ROI analysis was performed on the 6 volunteers, and the median stiffness values were compared between SE-EPI-MRE and TURBINE-MRE. RESULTS: The 3D wave-field images and the generated stiffness maps were comparable between TURBINE-MRE and standard SE-EPI-MRE for the phantom and the volunteers. The Bland-Altman plot showed no significant difference in the median regional stiffness values between the two methods. The stiffness measured with the 2 methods had a strong linear relationship with a Pearson correlation coefficient of 0.943. CONCLUSION: We demonstrated the feasibility of the new TURBINE-MRE sequence for acquiring the desired 3D wave-field data and stiffness maps in a phantom and in-vivo brains. This pilot study encourages further exploration of TURBINE-MRE for functional MRE, free-breathing abdominal MRE, and cardiac MRE applications.
Asunto(s)
Palabras clave

Texto completo: 1 Colección: 01-internacional Banco de datos: MEDLINE Asunto principal: Diagnóstico por Imagen de Elasticidad Tipo de estudio: Diagnostic_studies Límite: Humans Idioma: En Revista: Magn Reson Med Asunto de la revista: DIAGNOSTICO POR IMAGEM Año: 2021 Tipo del documento: Article País de afiliación: Estados Unidos

Texto completo: 1 Colección: 01-internacional Banco de datos: MEDLINE Asunto principal: Diagnóstico por Imagen de Elasticidad Tipo de estudio: Diagnostic_studies Límite: Humans Idioma: En Revista: Magn Reson Med Asunto de la revista: DIAGNOSTICO POR IMAGEM Año: 2021 Tipo del documento: Article País de afiliación: Estados Unidos