Microdosimetry of a therapeutic proton beam with a mini-TEPC and a MicroPlus-Bridge detector for RBE assessment.
Conte, V; Agosteo, S; Bianchi, A; Bolst, D; Bortot, D; Catalano, R; Cirrone, G A P; Colautti, P; Cuttone, G; Guatelli, S; James, B; Mazzucconi, D; Rosenfeld, A B; Selva, A; Tran, L; Petringa, G.
Afiliación
  • Conte V; INFN Laboratori Nazionali di Legnaro, viale dell'Università 2 35020 Legnaro, Italy.
  • Agosteo S; Politecnico di Milano, Dipartimento di Energia, via La Masa 34, Milano, Italy.
  • Bianchi A; Istituto Nazionale di Fisica Nucleare INFN - Sezione di Milano, via Celoria 16, Milano, Italy.
  • Bolst D; INFN Laboratori Nazionali di Legnaro, viale dell'Università 2 35020 Legnaro, Italy.
  • Bortot D; Belgian Nuclear Research Centre, SCK•CEN, Boeretang 200 2400 Mol, Belgium.
  • Catalano R; UHasselt, Faculty of Engineering Technology, Centre for Environmental Sciences, Nuclear Technology Center, Agoralaan 3590 Diepenbeek, Belgium.
  • Cirrone GAP; Centre for Medical Radiation Physics, University of Wollongong, Wollongong NSW 2522, Australia.
  • Colautti P; Politecnico di Milano, Dipartimento di Energia, via La Masa 34, Milano, Italy.
  • Cuttone G; Istituto Nazionale di Fisica Nucleare INFN - Sezione di Milano, via Celoria 16, Milano, Italy.
  • Guatelli S; INFN Laboratori Nazionali del Sud, via S.Sofia 62 95123 Catania, Italy.
  • James B; INFN Laboratori Nazionali del Sud, via S.Sofia 62 95123 Catania, Italy.
  • Mazzucconi D; INFN Laboratori Nazionali di Legnaro, viale dell'Università 2 35020 Legnaro, Italy.
  • Rosenfeld AB; INFN Laboratori Nazionali del Sud, via S.Sofia 62 95123 Catania, Italy.
  • Selva A; Centre for Medical Radiation Physics, University of Wollongong, Wollongong NSW 2522, Australia.
  • Tran L; Centre for Medical Radiation Physics, University of Wollongong, Wollongong NSW 2522, Australia.
  • Petringa G; Politecnico di Milano, Dipartimento di Energia, via La Masa 34, Milano, Italy.
Phys Med Biol ; 65(24): 245018, 2020 12 22.
Article en En | MEDLINE | ID: mdl-33086208
ABSTRACT
Proton beams are widely used worldwide to treat localized tumours, the lower entrance dose and no exit dose, thus sparing surrounding normal tissues, being the main advantage of this treatment modality compared to conventional photon techniques. Clinical proton beam therapy treatment planning is based on the use of a general relative biological effectiveness (RBE) of 1.1 along the whole beam penetration depth, without taking into account the documented increase in RBE at the end of the depth dose profile, in the Bragg peak and beyond. However, an inaccurate estimation of the RBE can cause both underdose or overdose, in particular it can cause the unfavourable situation of underdosing the tumour and overdosing the normal tissue just beyond the tumour, which limits the treatment success and increases the risk of complications. In view of a more precise dose delivery that takes into account the variation of RBE, experimental microdosimetry offers valuable tools for the quality assurance of LET or RBE-based treatment planning systems. The purpose of this work is to compare the response of two different microdosimetry systems the mini-TEPC and the MicroPlus-Bridge detector. Microdosimetric spectra were measured across the 62 MeV spread out Bragg peak of CATANA with the mini-TEPC and with the Bridge microdosimeter. The frequency and dose distributions of lineal energy were compared and the different contributions to the spectra were analysed, discussing the effects of different site sizes and chord length distributions. The shape of the lineal energy distributions measured with the two detectors are markedly different, due to the different water-equivalent sizes of the sensitive volumes 0.85 µm for the TEPC and 17.3 µm for the silicon detector. When the Loncol's biological weighting function is applied to calculate the microdosimetric assessment of the RBE, both detectors lead to results that are consistent with biological survival data for glioma U87 cells. Both the mini-TEPC and the MicroPlus-Bridge detector can be used to assess the RBE variation of a 62 MeV modulated proton beam along its penetration depth. The microdosimetric assessment of the RBE based on the Loncol's weighting function is in good agreement with radiobiological results when the 10% biological uncertainty is taken into account.
Asunto(s)

Texto completo: 1 Colección: 01-internacional Banco de datos: MEDLINE Asunto principal: Radiometría / Efectividad Biológica Relativa / Terapia de Protones Límite: Humans Idioma: En Revista: Phys Med Biol Año: 2020 Tipo del documento: Article País de afiliación: Italia

Texto completo: 1 Colección: 01-internacional Banco de datos: MEDLINE Asunto principal: Radiometría / Efectividad Biológica Relativa / Terapia de Protones Límite: Humans Idioma: En Revista: Phys Med Biol Año: 2020 Tipo del documento: Article País de afiliación: Italia