Performance of the Emprint and Amica Microwave Ablation Systems in ex vivo Porcine Livers: Sphericity and Reproducibility Versus Size.
Hendriks, P; Berkhout, W E M; Kaanen, C I; Sluijter, J H; Visser, I J; van den Dobbelsteen, J J; de Geus-Oei, L F; Webb, A G; Burgmans, M C.
Afiliación
  • Hendriks P; Department of Radiology, Leiden University Medical Center, P.O. Box 9600, RC Leiden, The Netherlands. p.hendriks@lumc.nl.
  • Berkhout WEM; Department of Radiology, Leiden University Medical Center, P.O. Box 9600, RC Leiden, The Netherlands.
  • Kaanen CI; Department of Radiology, Leiden University Medical Center, P.O. Box 9600, RC Leiden, The Netherlands.
  • Sluijter JH; Department of Radiology, Leiden University Medical Center, P.O. Box 9600, RC Leiden, The Netherlands.
  • Visser IJ; Department of Radiology, Leiden University Medical Center, P.O. Box 9600, RC Leiden, The Netherlands.
  • van den Dobbelsteen JJ; Medical Instruments & Bio-Inspired Technology, Faculty of Mechanical Engineering, Delft University of Technology, Delft, The Netherlands.
  • de Geus-Oei LF; Department of Radiology, Leiden University Medical Center, P.O. Box 9600, RC Leiden, The Netherlands.
  • Webb AG; Biomedical Photonic Imaging Group, University of Twente, Enschede, The Netherlands.
  • Burgmans MC; C. J. Gorter Center for High Field MRI, Department of Radiology, Leiden University Medical Center, Leiden, The Netherlands.
Cardiovasc Intervent Radiol ; 44(6): 952-958, 2021 Jun.
Article en En | MEDLINE | ID: mdl-33462682
ABSTRACT

PURPOSE:

To investigate the performance of two microwave ablation (MWA) systems regarding ablation volume, ablation shape and variability. MATERIALS AND

METHODS:

In this ex vivo study, the Emprint and Amica MWA systems were used to ablate porcine livers at 4 different settings of time and power (3 and 5 minutes at 60 and 80 Watt). In total, 48 ablations were analysed for ablation size and shape using Vitrea Advanced Visualization software after acquisition of a 7T MRI scan.

RESULTS:

Emprint ablations were smaller (11,1 vs. 21,1 mL p < 0.001), more spherical (sphericity index of 0.89 vs. 0.59 p < 0.001) and showed less variability than Amica ablations. In both systems, longer ablation time and higher power resulted in significantly larger ablation volumes.

CONCLUSION:

Emprint ablations were more spherical, and the results showed a lower variability than those of Amica ablations. This comes at the price of smaller ablation volumes.