Identification of deliberate catheter motion at the left atrial posterior wall during pulmonary vein isolation: Validity of respiratory motion adjustment.
Tomlinson, David R; Biscombe, Katie; True, John; Hosking, Joanne; Streeter, Adam J.
Afiliación
  • Tomlinson DR; South West Cardiothoracic Centre, Derriford Hospital, University Hospitals Plymouth NHS Trust, Plymouth, UK.
  • Biscombe K; Department of Medical Statistics, University of Plymouth Faculty of Health and Human Sciences, Plymouth, UK.
  • True J; Department of Medical Statistics, University of Plymouth Faculty of Health and Human Sciences, Plymouth, UK.
  • Hosking J; Department of Medical Statistics, University of Plymouth Faculty of Health and Human Sciences, Plymouth, UK.
  • Streeter AJ; Department of Medical Statistics, University of Plymouth Faculty of Health and Human Sciences, Plymouth, UK.
J Cardiovasc Electrophysiol ; 32(4): 994-1004, 2021 04.
Article en En | MEDLINE | ID: mdl-33570226
ABSTRACT

BACKGROUND:

During automated radiofrequency (RF) annotation-guided pulmonary vein isolation (PVI), respiratory motion adjustment (RMA) is recommended, yet lacks in vivo validation.

METHODS:

Following contact force (CF) PVI (continuous RF, 30 W) using general anesthesia and automated RF annotation-guidance (VISITAG™ force-over-time 100% minimum 1 g; 2 mm position stability; ACCURESP™ RMA "off") in 25 patients, we retrospectively examined RMA settings "on" versus "off" at the left atrial posterior wall (LAPW).

RESULTS:

Respiratory motion detection occurred in eight, permitting offline retrospective comparison of RMA settings. Significant differences in LAPW RF auto-annotation occurred according to RMA setting, with curves displaying catheter position, CF and impedance data indicating "best-fit" for catheter motion detection using RMA "off." Comparing RMA "on" versus "off," respectively total annotated sites, 82 versus 98; median RF duration per-site, 13.3 versus 10.6 s (p < 0.0001); median force time integral 177 versus 130 gs (p = 0.0002); mean inter-tag distance (ITD), 6.0 versus 4.8 mm (p = 0.002). Considering LAPW annotated site 1-to-2 transitions resulting from deliberate catheter movement, 3 concurrent with inadvertent 0 g CF demonstrated < 0.6 s difference in RF duration. However, 13 deliberate catheter movements during constant tissue contact (ITD range 2.1-7.0 mm) demonstrated (mean) site-1 RF duration difference 3.7 s (range -1.3 to 11.3 s) considering multiple measures of catheter position instability, the appropriate indication of deliberate catheter motion occurred with RMA "off" in all.

CONCLUSIONS:

ACCURESP™ respiratory motion adjustment importantly delayed the identification of deliberate and clinically relevant catheter motion during LAPW RF delivery, rendering auto-annotated RF display invalid. Operators seeking greater accuracy during auto-annotated RF delivery should avoid RMA use.
Asunto(s)
Palabras clave

Texto completo: 1 Colección: 01-internacional Banco de datos: MEDLINE Asunto principal: Venas Pulmonares / Fibrilación Atrial / Ablación por Catéter Tipo de estudio: Diagnostic_studies / Guideline / Observational_studies / Prognostic_studies Límite: Humans Idioma: En Revista: J Cardiovasc Electrophysiol Asunto de la revista: ANGIOLOGIA / CARDIOLOGIA / FISIOLOGIA Año: 2021 Tipo del documento: Article País de afiliación: Reino Unido

Texto completo: 1 Colección: 01-internacional Banco de datos: MEDLINE Asunto principal: Venas Pulmonares / Fibrilación Atrial / Ablación por Catéter Tipo de estudio: Diagnostic_studies / Guideline / Observational_studies / Prognostic_studies Límite: Humans Idioma: En Revista: J Cardiovasc Electrophysiol Asunto de la revista: ANGIOLOGIA / CARDIOLOGIA / FISIOLOGIA Año: 2021 Tipo del documento: Article País de afiliación: Reino Unido