Biomechanical optimization of the far cortical locking technique for early healing of distal femur fractures.
Sarwar, Ahmed; Gee, Aaron; Bougherara, Habiba; Kuzyk, Paul R T; Schemitsch, Emil H; Zdero, Radovan.
Afiliación
  • Sarwar A; Department of Mechanical and Industrial Engineering, Ryerson University, Toronto, ON, Canada.
  • Gee A; Orthopaedic Biomechanics Lab, Victoria Hospital, (Room A6-144), 800 Commissioners Road, London, ON N6A-5W9, Canada.
  • Bougherara H; Department of Mechanical and Industrial Engineering, Ryerson University, Toronto, ON, Canada.
  • Kuzyk PRT; Department of Surgery, University of Toronto, Toronto, ON, Canada.
  • Schemitsch EH; Orthopaedic Biomechanics Lab, Victoria Hospital, (Room A6-144), 800 Commissioners Road, London, ON N6A-5W9, Canada; Department of Surgery, Western University, London, ON, Canada.
  • Zdero R; Department of Mechanical and Industrial Engineering, Ryerson University, Toronto, ON, Canada; Orthopaedic Biomechanics Lab, Victoria Hospital, (Room A6-144), 800 Commissioners Road, London, ON N6A-5W9, Canada; Department of Surgery, Western University, London, ON, Canada; Department of Mechanical an
Med Eng Phys ; 89: 63-72, 2021 03.
Article en En | MEDLINE | ID: mdl-33608126
ABSTRACT
This finite element study optimized far cortical locking (FCL) technology for early callus formation in distal femur fracture fixation with a 9-hole plate using FCL screws proximal to, and standard locking screws distal to, the fracture. Analyses were done for 120 possible FCL screw configurations by varying FCL screw distribution and number. A hip joint force of 700 N (i.e. 100% x body weight) was used, which corresponds to a typical 140 N "toe-touch" foot-to-ground force (i.e. 20% x body weight) suggested to patients immediately after surgery. Increased FCL screw distribution (i.e. shorter plate working length) caused a decrease at the medial side and an increase at the lateral side of the axial interfragmentary motion (AIM), mildly affected shaft and condylar cortex Von Mises max stress (σMAX), increased plate σMAX, and decreased shaft FCL screw and condylar locking screw σMAX. Increased FCL screw number decreased AIM and σMAX on the shaft cortex, condylar cortex, plate, and FCL screws, but not condylar screws. The optimal FCL screw configuration had 3 FCL screws in plate holes #1, 5, and 6 (proximal to distal) for optimal AIM of 0.2 - 1 mm and reduce shear fracture motion, thereby encouraging early callus formation.
Asunto(s)
Palabras clave

Texto completo: 1 Colección: 01-internacional Banco de datos: MEDLINE Asunto principal: Tornillos Óseos / Fracturas del Fémur Límite: Humans Idioma: En Revista: Med Eng Phys Asunto de la revista: BIOFISICA / ENGENHARIA BIOMEDICA Año: 2021 Tipo del documento: Article País de afiliación: Canadá