Evaluation of interference from hemoglobin C, D, E and S traits on measurements of hemoglobin A1c by fifteen methods.
Rohlfing, Curt; Hanson, Steven; Estey, Mathew P; Bordeleau, Pierre; Little, Randie R.
Afiliación
  • Rohlfing C; Department of Pathology and Anatomical Sciences, University of Missouri, Columbia, MO, USA. Electronic address: rohlfingc@missouri.edu.
  • Hanson S; Department of Pathology and Anatomical Sciences, University of Missouri, Columbia, MO, USA.
  • Estey MP; DynaLIFE Medical Labs, Edmonton, Alberta, Canada.
  • Bordeleau P; DynaLIFE Medical Labs, Edmonton, Alberta, Canada.
  • Little RR; Department of Pathology and Anatomical Sciences, University of Missouri, Columbia, MO, USA.
Clin Chim Acta ; 522: 31-35, 2021 Nov.
Article en En | MEDLINE | ID: mdl-34352282
BACKGROUND: Hemoglobin C, D Punjab, E or S trait can interfere with hemoglobin A1c (HbA1c) results. We assessed whether they affect results obtained with 15 current assay methods. METHODS: Hemoglobin AA (HbAA), HbAC, HbAD Punjab, HbAE and HbAS samples were analyzed on 2 enzymatic, 4 ion-exchange HPLC and 9 immunoassay methods. Trinity Premier Hb9210 boronate affinity HPLC was the comparative method. An overall test of coincidence of least-squared linear regression lines was performed to determine if HbA1c results were statistically significantly different from those of HbAA samples. Clinically significant interference was defined as >6% difference from HbAA at 6 or 9% HbA1c compared to Premier Hb9210 using Deming regression. RESULTS: All methods showed statistically significant effects for one or more variants. Clinically significant effects were observed for the Tosoh G11 variant mode (HbAD), Roche b 101 (HbAC and HbAE) and Siemens DCA Vantage (HbAE and HbAS). All other methods (Beckman Coulter B93009 and B00389 on DxC700AU, and Unicel DxC, Ortho Clinical Vitros 5.1, Roche cobas c 513, Siemens Dimension RxL and Vista, and Enzymatic on Advia and Atellica, Tosoh G8 5.24 and 5.28, and GX) showed no clinically significant differences. CONCLUSIONS: A few methods showed interference from one or more variants. Laboratories need to be aware of potential HbA1c assay interferences.
Asunto(s)
Palabras clave

Texto completo: 1 Colección: 01-internacional Banco de datos: MEDLINE Asunto principal: Hemoglobina C / Pruebas Hematológicas Límite: Humans Idioma: En Revista: Clin Chim Acta Año: 2021 Tipo del documento: Article

Texto completo: 1 Colección: 01-internacional Banco de datos: MEDLINE Asunto principal: Hemoglobina C / Pruebas Hematológicas Límite: Humans Idioma: En Revista: Clin Chim Acta Año: 2021 Tipo del documento: Article