Correlation and agreement between superb micro-vascular imaging and contrast-enhanced ultrasound for assessing radiofrequency ablation treatment of thyroid nodules: a preliminary study.
Lan, Yu; Li, Nan; Song, Qing; Zhang, Ming-Bo; Luo, Yu-Kun; Zhang, Yan.
Afiliación
  • Lan Y; Department of Ultrasound, General Hospital of Chinese PLA, 28 Fuxing Road, Haidian District, Beijing, China.
  • Li N; School of Medicine, Nankai University, 94 Weijin Road, Nankai District, Tianjin, China.
  • Song Q; Department of Ultrasound, The People's Hospital of Liaoning Province, Shenyang, China.
  • Zhang MB; Department of Ultrasound, General Hospital of Chinese PLA, 28 Fuxing Road, Haidian District, Beijing, China.
  • Luo YK; Department of Ultrasound, General Hospital of Chinese PLA, 28 Fuxing Road, Haidian District, Beijing, China.
  • Zhang Y; Department of Ultrasound, General Hospital of Chinese PLA, 28 Fuxing Road, Haidian District, Beijing, China. owsifanduizhe@126.com.
BMC Med Imaging ; 21(1): 175, 2021 11 22.
Article en En | MEDLINE | ID: mdl-34809604
ABSTRACT

BACKGROUND:

To evaluate the correlation and agreement between superb micro-vascular imaging (SMI) mode and the contrast-enhanced ultrasound (CEUS) mode for the ablative completeness and the volumes of ablation lesions to determine the clinical application value of SMI in follow-up after radiofrequency ablation.

METHODS:

From April 2020 to June 2020, two radiologists used SMI and CEUS mode to measure the volume of the ablation lesion. We use intra-class correlation coefficient (ICC), scatter plots and Bland-Altman plots to evaluate the correlation and agreement of the two techniques. In addition, intra- and inter-observer reliability in volume measurement of ablation lesions with SMI mode was assessed.

RESULTS:

SMI mode and CEUS mode have good agreement in the evaluation of ablative completeness. The ICC was 0.876 and 0.928 of reader A and reader B between SMI mode and CEUS mode in terms of ablation lesions volume measurement. There was a strong correlation between the two modes in both reader A and reader B (rA = 0.808; rB = 0.882). The ICC was 0.836 for the inter-observer reliability of SMI technique. The scatter plot showed a good linear relation (r = 0.715). In the Bland-Altman plot, 4.35% (1/23) of the points was outside the 95% limits of agreement. The ICC was 0.965 for the intra-observer reliability of SMI technique, the scatter plot also showed a strong linear correlation (r = 0.965). In the Bland-Altman plot, 8.70% (2/23) of the points was outside the 95% limits of agreement.

CONCLUSIONS:

SMI and CEUS have good agreement and correlation in the ablation volume measurement. SMI technology is expected to be applied as an alternative to CEUS in the clinical follow-up of ablation lesions.
Asunto(s)
Palabras clave

Texto completo: 1 Colección: 01-internacional Banco de datos: MEDLINE Asunto principal: Ultrasonografía / Nódulo Tiroideo / Ablación por Radiofrecuencia Tipo de estudio: Diagnostic_studies Límite: Adult / Female / Humans / Male / Middle aged Idioma: En Revista: BMC Med Imaging Asunto de la revista: DIAGNOSTICO POR IMAGEM Año: 2021 Tipo del documento: Article País de afiliación: China

Texto completo: 1 Colección: 01-internacional Banco de datos: MEDLINE Asunto principal: Ultrasonografía / Nódulo Tiroideo / Ablación por Radiofrecuencia Tipo de estudio: Diagnostic_studies Límite: Adult / Female / Humans / Male / Middle aged Idioma: En Revista: BMC Med Imaging Asunto de la revista: DIAGNOSTICO POR IMAGEM Año: 2021 Tipo del documento: Article País de afiliación: China