Three-dimensional reduced field-of-view imaging (3D-rFOVI).
Sun, Kaibao; Zhong, Zheng; Dan, Guangyu; Karaman, Muge; Luo, Qingfei; Zhou, Xiaohong Joe.
Afiliación
  • Sun K; Center for Magnetic Resonance Research, University of Illinois at Chicago, Chicago, Illinois, USA.
  • Zhong Z; Center for Magnetic Resonance Research, University of Illinois at Chicago, Chicago, Illinois, USA.
  • Dan G; Department of Biomedical Engineering, University of Illinois at Chicago, Chicago, Illinois, USA.
  • Karaman M; Center for Magnetic Resonance Research, University of Illinois at Chicago, Chicago, Illinois, USA.
  • Luo Q; Department of Biomedical Engineering, University of Illinois at Chicago, Chicago, Illinois, USA.
  • Zhou XJ; Center for Magnetic Resonance Research, University of Illinois at Chicago, Chicago, Illinois, USA.
Magn Reson Med ; 87(5): 2372-2379, 2022 05.
Article en En | MEDLINE | ID: mdl-34894639
PURPOSE: This study aimed at developing a 3D reduced field-of-view imaging (3D-rFOVI) technique using a 2D radiofrequency (RF) pulse, and demonstrating its ability to achieve isotropic high spatial resolution and reduced image distortion in echo planar imaging (EPI). METHODS: The proposed 3D-rFOVI technique takes advantage of a 2D RF pulse to excite a slab along the conventional slice-selection direction (i.e., z-direction) while limiting the spatial extent along the phase-encoded direction (i.e., y-direction) within the slab. The slab is phase-encoded in both through-slab and in-slab phase-encoded directions. The 3D-rFOVI technique was implemented at 3T in gradient-echo and spin-echo EPI pulse sequences for functional MRI (fMRI) and diffusion-weighted imaging (DWI), respectively. 3D-rFOVI experiments were performed on a phantom and human brain to illustrate image distortion reduction, as well as isotropic high spatial resolution, in comparison with 3D full-FOV imaging. RESULTS: In both the phantom and the human brain, image voxel dislocation was substantially reduced by 3D-rFOVI when compared with full-FOV imaging. In the fMRI experiment with visual stimulation, 3D isotropic spatial resolution of (2 × 2 × 2 mm3 ) was achieved with an adequate signal-to-noise ratio (81.5) and blood oxygen level-dependent (BOLD) contrast (2.5%). In the DWI experiment, diffusion-weighted brain images with an isotropic resolution of (1 × 1 × 1 mm3 ) was obtained without appreciable image distortion. CONCLUSION: This study indicates that 3D-rFOVI is a viable approach to 3D neuroimaging over a zoomed region.
Asunto(s)
Palabras clave

Texto completo: 1 Colección: 01-internacional Banco de datos: MEDLINE Asunto principal: Imagen Eco-Planar / Imagen de Difusión por Resonancia Magnética Límite: Humans Idioma: En Revista: Magn Reson Med Asunto de la revista: DIAGNOSTICO POR IMAGEM Año: 2022 Tipo del documento: Article País de afiliación: Estados Unidos

Texto completo: 1 Colección: 01-internacional Banco de datos: MEDLINE Asunto principal: Imagen Eco-Planar / Imagen de Difusión por Resonancia Magnética Límite: Humans Idioma: En Revista: Magn Reson Med Asunto de la revista: DIAGNOSTICO POR IMAGEM Año: 2022 Tipo del documento: Article País de afiliación: Estados Unidos