Ganglion Cell Layer Thickness Variance Using SPECTRALIS Optical Coherence Tomography.
Mallory, Paul; Wagley, Sushant; Chen, John; McClelland, Collin M; Downie, Peter A; Lindgren, Bruce; Lee, Michael S.
Afiliación
  • Mallory P; Department of Ophthalmology and Visual Neuroscience (PM, SW, CMM, PAD, MSL), University of Minnesota, Minneapolis, Minnesota; Departments of Ophthalmology and Neurology (JC), Mayo Clinic, Rochester, Minnesota; and Clinical and Translational Science Institute (BL), University of Minnesota, Minneapolis, Minnesota.
J Neuroophthalmol ; 42(3): 310-315, 2022 09 01.
Article en En | MEDLINE | ID: mdl-35427252
ABSTRACT

BACKGROUND:

To determine the normal variance of the mean macular ganglion cell layer (GCL) volume among subjects without significant ocular pathology using SPECTRALIS optical coherence tomography (OCT).

METHODS:

Fifty subjects underwent a baseline scan using SPECTRALIS OCT followed by 2 more studies with (reg-ON) and without (reg-OFF) eye registration all taken at the same session. The mean GCL volume was measured using built-in SPECTRALIS software. Eyes with macular pathology were excluded. The reproducibility of the measurements of the GCL volume was evaluated with Bland-Altman plots and limits of agreement, intraclass correlation coefficient (ICC), and the coefficient of repeatability (CR).

RESULTS:

A total of 98 eyes met criteria for the analysis. The mean GCL volume difference was 0.0002 ± 0.029 and -0.0005 ± 0.035 mm 3 for scans 1 versus 2 (baseline vs reg-ON) and 3 (baseline vs reg-OFF), respectively. The ICCs were 0.985 and 0.977 for the baseline vs reg-ON and reg-OFF groups. The CR for baseline vs reg-ON was 0.056 while CR for baseline vs reg-OFF was 0.069. Ninety percent of eyes fell within 0.04 mm 3 of test-retest reliability.

CONCLUSIONS:

Our model found a predictable threshold of 0.07 mm 3 or less for SPECTRALIS OCT mean GCL volume variance, which did not significantly change with eye registration in eyes without macular pathology. Clinicians may also consider a threshold of 0.04 mm 3 when determining stable vs progressive changes in mean GCL volume using this device.
Asunto(s)

Texto completo: 1 Colección: 01-internacional Banco de datos: MEDLINE Asunto principal: Células Ganglionares de la Retina / Tomografía de Coherencia Óptica Tipo de estudio: Prognostic_studies Límite: Humans Idioma: En Revista: J Neuroophthalmol Asunto de la revista: NEUROLOGIA / OFTALMOLOGIA Año: 2022 Tipo del documento: Article

Texto completo: 1 Colección: 01-internacional Banco de datos: MEDLINE Asunto principal: Células Ganglionares de la Retina / Tomografía de Coherencia Óptica Tipo de estudio: Prognostic_studies Límite: Humans Idioma: En Revista: J Neuroophthalmol Asunto de la revista: NEUROLOGIA / OFTALMOLOGIA Año: 2022 Tipo del documento: Article