Your browser doesn't support javascript.
loading
Durability of left bundle branch area pacing.
Mehta, Nikhil A; Saqi, Bilal; Sabzwari, Syed R A; Gupta, Rahul; Vyas, Apurva; Gordon, Jeffrey; Cossu, Sergio; Nazir, Talha; Freudenberger, Ronald; Bozorgnia, Babak.
Afiliación
  • Mehta NA; Department of Electrophysiology, Lehigh Valley Heart Institute, Allentown, Pennsylvania, USA.
  • Gupta R; Department of Cardiology, Lehigh Valley Heart Institute, Allentown, Pennsylvania, USA.
  • Vyas A; Department of Cardiology, Lehigh Valley Heart Institute, Allentown, Pennsylvania, USA.
  • Gordon J; Department of Electrophysiology, Lehigh Valley Heart Institute, Allentown, Pennsylvania, USA.
  • Cossu S; Department of Electrophysiology, Lehigh Valley Heart Institute, Allentown, Pennsylvania, USA.
  • Nazir T; Department of Electrophysiology, Lehigh Valley Heart Institute, Allentown, Pennsylvania, USA.
  • Freudenberger R; Department of Cardiology, Lehigh Valley Heart Institute, Allentown, Pennsylvania, USA.
  • Bozorgnia B; Department of Electrophysiology, Lehigh Valley Heart Institute, Allentown, Pennsylvania, USA.
J Cardiovasc Electrophysiol ; 33(7): 1529-1536, 2022 07.
Article en En | MEDLINE | ID: mdl-35586896
ABSTRACT

BACKGROUND:

Left bundle branch area pacing (LBBAP) is a form of conduction system pacing. Long-term data on the safety and performance of LBBAP 1 year postdevice implantation has not been well described. METHODS AND

RESULTS:

Sixty-five patients (49% females) who received LBBAP for bradycardia indications using the SelectSecure 3830 lead (Medtronic) were retrospectively evaluated. Clinical variables were examined. Lead parameters were obtained at implant and during regular follow-up. Mean age of patients was 75.7 ± 10.1 years with left ventricular ejection fraction 59.8 ± 10.4%. Indications for pacing were atrioventricular block 55%, sinus node dysfunction 19%, tachy-brady syndrome 15%, atrioventricular node ablation 8%, and bail out cardiac resynchronization therapy 3%. Mean baseline QRS measured 120 ± 38 ms, paced QRS duration was 138 ± 22ms. Paced QRS narrowed by 24 ms in those with pre-existing left bundle branch block (BBB), increased by 1 ms in those with pre-existing right BBB, and increased by 42 ms in those with no BBB. LBBAP threshold at implant was 0.521 ± 0.153 V at 0.4 ms, and increased to 0.654 ± 0.186 V at 3 months (+26%), 0.707 ± 0.186 V at 6 months (+36%), and 0.772 ± 0.220 V at 12 months (+48%). Patients with left BBB showed the maximum benefit with QRS narrowing 24 ms. Pacing impedance remained unchanged with no procedure-related complications.

CONCLUSION:

LBBAP is a durable form of conduction system pacing with pacing thresholds remaining relatively stable over 12 months post device implantation. Patients with left BBB display the narrowest paced QRS.
Asunto(s)
Palabras clave

Texto completo: 1 Colección: 01-internacional Banco de datos: MEDLINE Asunto principal: Fascículo Atrioventricular / Terapia de Resincronización Cardíaca Tipo de estudio: Observational_studies Límite: Aged / Aged80 / Female / Humans / Male Idioma: En Revista: J Cardiovasc Electrophysiol Asunto de la revista: ANGIOLOGIA / CARDIOLOGIA / FISIOLOGIA Año: 2022 Tipo del documento: Article País de afiliación: Estados Unidos

Texto completo: 1 Colección: 01-internacional Banco de datos: MEDLINE Asunto principal: Fascículo Atrioventricular / Terapia de Resincronización Cardíaca Tipo de estudio: Observational_studies Límite: Aged / Aged80 / Female / Humans / Male Idioma: En Revista: J Cardiovasc Electrophysiol Asunto de la revista: ANGIOLOGIA / CARDIOLOGIA / FISIOLOGIA Año: 2022 Tipo del documento: Article País de afiliación: Estados Unidos