Your browser doesn't support javascript.
loading
Leadless pacemaker in Twiddler's syndrome: A case report.
Ducceschi, Valentino; De Divitiis, Marcello; Ciriello, Giovanni Domenico; Gallo, Paolo.
Afiliación
  • Ducceschi V; Electrophysiology and Cardiac Pacing Unit, Pellegrini Hospital, Naples, Italy.
  • De Divitiis M; Electrophysiology and Cardiac Pacing Unit, Pellegrini Hospital, Naples, Italy.
  • Ciriello GD; Electrophysiology and Cardiac Pacing Unit, Pellegrini Hospital, Naples, Italy.
  • Gallo P; Cardiology Unit, C.D.C. Villa dei Fiori, Acerra, Napoli, Italy.
Pacing Clin Electrophysiol ; 45(9): 1062-1064, 2022 09.
Article en En | MEDLINE | ID: mdl-35906933
ABSTRACT
Twiddler's syndrome is a rare cause of pacemaker lead dislodgement. We present the case of a 49-year-old male patient with Down's syndrome implanted with a dual chamber pacemaker showing high ventricular impedance, no sensing, and complete loss of capture for both leads at the 3-month follow-up due to Twiddler Syndrome. The dislocated device was removed, and an endocardial leadless pacing system was implanted.
Asunto(s)
Palabras clave

Texto completo: 1 Colección: 01-internacional Banco de datos: MEDLINE Asunto principal: Marcapaso Artificial Límite: Humans / Male / Middle aged Idioma: En Revista: Pacing Clin Electrophysiol Año: 2022 Tipo del documento: Article País de afiliación: Italia

Texto completo: 1 Colección: 01-internacional Banco de datos: MEDLINE Asunto principal: Marcapaso Artificial Límite: Humans / Male / Middle aged Idioma: En Revista: Pacing Clin Electrophysiol Año: 2022 Tipo del documento: Article País de afiliación: Italia