Assessment of a new KoMaWo electrode-patch configuration accuracy and review of the literature.
Boszko, Maria; Osak, Gabriela; Zurawska, Natalia; Skoczylas, Kamila; Krzowski, Bartosz; Wróblewski, Grzegorz; Maciejewski, Adrian; Sobiech, Judyta; Ostrowski, Szymon; Grabowski, Marcin; Koltowski, Lukasz.
Afiliación
  • Boszko M; 1st Chair and Department of Cardiology, Medical University of Warsaw, Warsaw, Poland.
  • Osak G; 1st Chair and Department of Cardiology, Medical University of Warsaw, Warsaw, Poland.
  • Zurawska N; 1st Chair and Department of Cardiology, Medical University of Warsaw, Warsaw, Poland.
  • Skoczylas K; 1st Chair and Department of Cardiology, Medical University of Warsaw, Warsaw, Poland.
  • Krzowski B; 1st Chair and Department of Cardiology, Medical University of Warsaw, Warsaw, Poland. Electronic address: bartekkrzowski@gmail.com.
  • Wróblewski G; Institute of Metrology and Biomedical Engineering, Warsaw University of Technology, Warsaw, Poland.
  • Maciejewski A; 1st Chair and Department of Cardiology, Medical University of Warsaw, Warsaw, Poland.
  • Sobiech J; Faculty of Physics, Warsaw University of Technology, Warsaw, Poland.
  • Ostrowski S; Institute of Metrology and Biomedical Engineering, Warsaw University of Technology, Warsaw, Poland.
  • Grabowski M; 1st Chair and Department of Cardiology, Medical University of Warsaw, Warsaw, Poland.
  • Koltowski L; 1st Chair and Department of Cardiology, Medical University of Warsaw, Warsaw, Poland.
J Electrocardiol ; 75: 82-87, 2022.
Article en En | MEDLINE | ID: mdl-35918203
INTRODUCTION: Standard 12­lead electrocardiogram (ECG) is a basic element of routine everyday clinical practice. Traditional cardiac monitoring devices are associated with considerable limitations. Adhesive patches, novel digital solutions, may become a useful diagnostic tool for several cardiovascular diseases. MATERIALS AND METHODS: We propose a new variation of ECG electrodes positioning called KoMaWo. 15 consecutive patients presenting with ST segment deviations due to coronary artery disease were enrolled. The accuracy and utility of the new configuration was assessed and compared with the Mason-Likar configuration, as well as with a standard 12­lead ECG recording. The scans were blinded and interpreted by two independent cardiologists. RESULTS: There were no statistically significant differences in morphology, as well as in the duration of individual waves, complexes, segments, and intervals between the scans obtained using all three methods. In a subgroup analysis, with regard to age, body mass and left ventricle ejection fraction (LVEF), KoMaWo was non-inferior to standard ECG with a 0.2 mm margin. DISCUSSION: The role of traditional cardiac monitoring devices is recognized as the gold standard of patient management. However, certain limitations should be considered. Adhesive patches are light-weight, well-tolerated and do not interfere with daily activities of patients. These novel devices allow for extended monitoring, facilitating increased diagnostic accuracy, regarding cardiac arrhythmias. CONCLUSIONS: The KoMaWo configuration is not inferior to standard electrode placement, nor to Mason-Likar configuration, including its ability to capture ST segment deviations. Adhesive patches may become a valid alternative for traditional cardiac monitoring methods.