ADEPT: Accurate Diffusion Echo-Planar imaging with multi-contrast shoTs.
Shafieizargar, Banafshe; Jeurissen, Ben; Poot, Dirk H J; Klein, Stefan; Van Audekerke, Johan; Verhoye, Marleen; den Dekker, Arnold J; Sijbers, Jan.
Afiliación
  • Shafieizargar B; imec-Vision Lab, Department of Physics, University of Antwerp, Antwerp, Belgium.
  • Jeurissen B; NEURO Research Centre of Excellence, University of Antwerp, Antwerp, Belgium.
  • Poot DHJ; imec-Vision Lab, Department of Physics, University of Antwerp, Antwerp, Belgium.
  • Klein S; NEURO Research Centre of Excellence, University of Antwerp, Antwerp, Belgium.
  • Van Audekerke J; Department of Radiology and Nuclear Medicine, Biomedical Imaging Group Rotterdam, Erasmus MC, Rotterdam, The Netherlands.
  • Verhoye M; Department of Radiology and Nuclear Medicine, Biomedical Imaging Group Rotterdam, Erasmus MC, Rotterdam, The Netherlands.
  • den Dekker AJ; NEURO Research Centre of Excellence, University of Antwerp, Antwerp, Belgium.
  • Sijbers J; Bio-Imaging Lab, Department of Biomedical Sciences, University of Antwerp, Antwerp, Belgium.
Magn Reson Med ; 89(1): 396-410, 2023 01.
Article en En | MEDLINE | ID: mdl-36110059
ABSTRACT

PURPOSE:

To introduce a novel imaging and parameter estimation framework for accurate multi-shot diffusion MRI. THEORY AND

METHODS:

We propose a new framework called ADEPT (Accurate Diffusion Echo-Planar imaging with multi-contrast shoTs) that enables fast diffusion MRI by allowing diffusion contrast settings to change between shots in a multi-shot EPI acquisition (i.e., intra-scan modulation). The framework estimates diffusion parameter maps directly from the acquired intra-scan modulated k-space data, while simultaneously accounting for shot-to-shot phase inconsistencies. The performance of the estimation framework is evaluated using Monte Carlo simulation studies and in-vivo experiments and compared to that of reference methods that rely on parallel imaging for shot-to-shot phase correction.

RESULTS:

Simulation and real-data experiments show that ADEPT yields more accurate and more precise estimates of the diffusion metrics in multi-shot EPI data in comparison with the reference methods.

CONCLUSION:

ADEPT allows fast multi-shot EPI diffusion MRI without significantly degrading the accuracy and precision of the estimated diffusion maps.
Asunto(s)
Palabras clave

Texto completo: 1 Colección: 01-internacional Banco de datos: MEDLINE Asunto principal: Procesamiento de Imagen Asistido por Computador / Imagen Eco-Planar Idioma: En Revista: Magn Reson Med Asunto de la revista: DIAGNOSTICO POR IMAGEM Año: 2023 Tipo del documento: Article País de afiliación: Bélgica

Texto completo: 1 Colección: 01-internacional Banco de datos: MEDLINE Asunto principal: Procesamiento de Imagen Asistido por Computador / Imagen Eco-Planar Idioma: En Revista: Magn Reson Med Asunto de la revista: DIAGNOSTICO POR IMAGEM Año: 2023 Tipo del documento: Article País de afiliación: Bélgica