Your browser doesn't support javascript.
loading
Outcomes of modular cable exchange in HeartMate 3 patients versus conventional driveline repair in HeartMate II patients.
Merzah, Ali Saad; Hanke, Jasmin S; Li, Tong; Saad Merzah, Fadi A; Gabriel, Maria; Derda, Anselm A; Homann, Katharina; Haverich, Axel; Schmitto, Jan D; Dogan, Günes.
Afiliación
  • Merzah AS; Department of Cardiac-, Thoracic-, Transplantation and Vascular Surgery, Hannover Medical School, Hannover, Germany.
  • Hanke JS; Department of Cardiac-, Thoracic-, Transplantation and Vascular Surgery, Hannover Medical School, Hannover, Germany.
  • Li T; Department of Cardiac-, Thoracic-, Transplantation and Vascular Surgery, Hannover Medical School, Hannover, Germany.
  • Saad Merzah FA; Department of Cardiac-, Thoracic-, Transplantation and Vascular Surgery, Hannover Medical School, Hannover, Germany.
  • Gabriel M; Department of Neurology, Hannover Medical School, Hannover, Germany.
  • Derda AA; Department of Cardiology and Angiology, Hannover Medical School, Hannover, Germany.
  • Homann K; Department of Cardiac-, Thoracic-, Transplantation and Vascular Surgery, Hannover Medical School, Hannover, Germany.
  • Haverich A; Department of Cardiac-, Thoracic-, Transplantation and Vascular Surgery, Hannover Medical School, Hannover, Germany.
  • Schmitto JD; Department of Cardiac-, Thoracic-, Transplantation and Vascular Surgery, Hannover Medical School, Hannover, Germany.
  • Dogan G; Department of Cardiac-, Thoracic-, Transplantation and Vascular Surgery, Hannover Medical School, Hannover, Germany.
Artif Organs ; 47(2): 380-386, 2023 Feb.
Article en En | MEDLINE | ID: mdl-36148849
ABSTRACT

BACKGROUND:

Driveline (DL) damages are a common difficulty among ventricular assist devices (VAD). Repairing the electrical fibers inside the DL on a running pump is hazardous and requires technical expertise, which is not easily available on site. A new feature of the HeartMate3 (Abbott, U.S.A.) LVAD is a modular driveline that allows an easy exchange of the DL cord. In this report we analyze our experiences with this feature.

METHODS:

We performed a retrospective analysis of 302 patients who underwent either HeartMate II or HeartMate 3 implantation between February 2004 and September 2021. Patients were screened for driveline faults and need for exchange or repair of driveline or VAD exchange. Documented were baseline characteristics, reasons for DL or VAD exchange, and complications. Follow-up was three months after the procedure.

RESULTS:

We present a cohort of 302 patients who underwent either HMII (n = 107;35.4%) or HM 3 (n = 195; 64.5%) implantation. Out of those, 40 patients (40/302; 13.2%) required driveline repair (DLRe) or exchange (DLEx). Out of 107 HMII patients, 9 showed severe DL damages (9/107; 8.4%). Six patients (6/9; 66.6%) underwent DLRe, two patients (2/6; 33.3%) required VAD exchange after DLRe, one patient (1/2;50%) experienced emergency VAD exchange after pump stop. The DLRe procedure in the other four patients (4/6; 66.6%) was successful. Due to damage to the internal driveline two patients (2/9; 22.2%) underwent emergency device exchange and one patient (1/9;11.1%) was listed for transplantation. 31 out 195 HM3 patients underwent exchange of the modular DL. In none of the cases, damages of the internal fibers were the reasons for the exchange. In 100% of the cases, damages of the external coating were the reason for DL exchange. In none of the cases, complications occurred after the exchange procedure.

CONCLUSIONS:

Driveline damages are a habitual, recurrent complication in VAD patients. The exchange of the modular driveline cable of the HM3 is feasible and safe compared to the conventional DL repair in HMII patients. Risky repair attempts and surgical LVAD exchange due to major damages of the electrical fibers can be avoided successfully by the new feature of HM3 driveline.