Your browser doesn't support javascript.
loading
Long-Term Valve Durability in Patients Undergoing Transcatheter Aortic Valve Implantation.
Stehli, Julia; Dagan, Misha; Duffy, Stephen J; Morisset, Stéphane; Vriesendorp, Pieter; Nanayakkara, Shane; Samuel Koh, Ji Quan; Quine, Edward; Htun, Nay M; Stub, Dion; Dick, Ron; Walton, Antony S.
Afiliación
  • Stehli J; Department of Cardiology, Alfred Hospital, Melbourne, Vic, Australia; Cardiovascular Institute, Epworth Healthcare, Melbourne, Vic, Australia; School of Clinical Sciences at Monash Health, Monash Cardiovascular Research Centre, Monash University, Melbourne, Vic, Australia.
  • Dagan M; Department of Cardiology, Alfred Hospital, Melbourne, Vic, Australia.
  • Duffy SJ; Department of Cardiology, Alfred Hospital, Melbourne, Vic, Australia; Centre of Cardiovascular Research and Education in Therapeutics, Department of Epidemiology and Preventive Medicine, Monash University, Melbourne, Vic, Australia; Baker IDI Heart and Diabetes Institute, Melbourne, Vic, Australia.
  • Morisset S; Statistics consultancy, Pérouges, France.
  • Vriesendorp P; Department of Cardiology, Alfred Hospital, Melbourne, Vic, Australia.
  • Nanayakkara S; Department of Cardiology, Alfred Hospital, Melbourne, Vic, Australia; Centre of Cardiovascular Research and Education in Therapeutics, Department of Epidemiology and Preventive Medicine, Monash University, Melbourne, Vic, Australia; Baker IDI Heart and Diabetes Institute, Melbourne, Vic, Australia.
  • Samuel Koh JQ; Department of Cardiology, Alfred Hospital, Melbourne, Vic, Australia.
  • Quine E; Department of Cardiology, Alfred Hospital, Melbourne, Vic, Australia.
  • Htun NM; Department of Cardiology, Alfred Hospital, Melbourne, Vic, Australia.
  • Stub D; Department of Cardiology, Alfred Hospital, Melbourne, Vic, Australia; Centre of Cardiovascular Research and Education in Therapeutics, Department of Epidemiology and Preventive Medicine, Monash University, Melbourne, Vic, Australia; Baker IDI Heart and Diabetes Institute, Melbourne, Vic, Australia.
  • Dick R; Cardiovascular Institute, Epworth Healthcare, Melbourne, Vic, Australia.
  • Walton AS; Department of Cardiology, Alfred Hospital, Melbourne, Vic, Australia; Cardiovascular Institute, Epworth Healthcare, Melbourne, Vic, Australia. Electronic address: a.walton@alfred.org.au.
Heart Lung Circ ; 32(2): 240-246, 2023 Feb.
Article en En | MEDLINE | ID: mdl-36376193
ABSTRACT

AIMS:

To evaluate the long-term incidence of structural valve deterioration (SVD) in patients who underwent transcatheter aortic valve implantation (TAVI). METHOD AND

RESULTS:

Between 2008 and 2018, 693 underwent TAVI at two centres. Four hundred and twenty-one (421) patients (mean age 83.6±6.0 yrs) survived for ≥2 years post TAVI and had at least two consecutive transthoracic echocardiographies (TTEs) with the latest TTE no less than 2 years after TAVI, and were therefore included in the analysis for SVD. Median follow-up was 4.7 (3.6-6.0) years and median echocardiography follow-up 3 (3.0-4.0) years. All-cause mortality was 30.9% (130) with a median time to death of 4.1 (3.0-5.6) years. The cumulative incidence of SVD increased from 1.7% (95% CI, 0.4-2.9) at 3 years to 3.5% (95% CI, 1.5-5.8) at 5 years and 4.7% (95% CI, 1.6-7.9) at 10 years. The overall median time to SVD was 3 (2-4) years. Twelve (12) patients demonstrated SVD stage 2, and 1 patient stage 3. No SVD required re-intervention. All other patients showed no significant changes in valve parameters over time.

CONCLUSIONS:

Structural valve deterioration is an uncommon event, occurring in 5% over a total follow-up of 10 years. Most patients show stable valve parameters. However, the analysis is limited by the loss of follow-up (owing to patient mortality), which renders extrapolation of the data to a younger patient population difficult.
Asunto(s)
Palabras clave

Texto completo: 1 Colección: 01-internacional Banco de datos: MEDLINE Asunto principal: Estenosis de la Válvula Aórtica / Prótesis Valvulares Cardíacas / Reemplazo de la Válvula Aórtica Transcatéter Límite: Aged / Aged80 / Humans Idioma: En Revista: Heart Lung Circ Asunto de la revista: ANGIOLOGIA / CARDIOLOGIA Año: 2023 Tipo del documento: Article País de afiliación: Australia

Texto completo: 1 Colección: 01-internacional Banco de datos: MEDLINE Asunto principal: Estenosis de la Válvula Aórtica / Prótesis Valvulares Cardíacas / Reemplazo de la Válvula Aórtica Transcatéter Límite: Aged / Aged80 / Humans Idioma: En Revista: Heart Lung Circ Asunto de la revista: ANGIOLOGIA / CARDIOLOGIA Año: 2023 Tipo del documento: Article País de afiliación: Australia