Preclinical evaluation of semi-automated laser ablation for pulmonary vein isolation: A comparative study.
Kuroki, Kenji; Reddy, Vivek Y; Iwasawa, Jin; Kawamura, Iwanari; Neuzil, Petr; Estabrook, Brian; Melsky, Gerald; Dukkipati, Srinivas R; Koruth, Jacob.
Afiliación
  • Kuroki K; Helmsley Electrophysiology Center, Icahn School of Medicine at Mount Sinai, New York, New York, USA.
  • Reddy VY; Helmsley Electrophysiology Center, Icahn School of Medicine at Mount Sinai, New York, New York, USA.
  • Iwasawa J; Helmsley Electrophysiology Center, Icahn School of Medicine at Mount Sinai, New York, New York, USA.
  • Kawamura I; Helmsley Electrophysiology Center, Icahn School of Medicine at Mount Sinai, New York, New York, USA.
  • Neuzil P; Department of Cardiology, Hospital Na Homolce, Prague, Czech Republic.
  • Estabrook B; CardioFocus, Inc, Marlborough, Massachusetts, USA.
  • Melsky G; CardioFocus, Inc, Marlborough, Massachusetts, USA.
  • Dukkipati SR; Helmsley Electrophysiology Center, Icahn School of Medicine at Mount Sinai, New York, New York, USA.
  • Koruth J; Helmsley Electrophysiology Center, Icahn School of Medicine at Mount Sinai, New York, New York, USA.
J Cardiovasc Electrophysiol ; 34(2): 315-324, 2023 02.
Article en En | MEDLINE | ID: mdl-36511472
ABSTRACT

INTRODUCTION:

Visually-guided laser balloon ablation (VGLA) currently requires careful manual rotation of the laser to create overlapping lesions. A novel semi-automated VGLA may reduce ablation times and lesion gaps. We aimed to compare semi-automated (SA) VGLA to that of manual (MN) VGLA.

METHODS:

Acute Nine swine underwent right superior pulmonary vein isolation (PVI) using either SA (n = 3, 13-18 W), MN (n = 3, 8.5-12 W), or radiofrequency (RF, n = 3, 25-40 W) and were killed acutely. Chronic 16 swine, underwent PVI using either SA (n = 8, 15 W) or MN (n = 8, 10 W), and were survived for 1 month before being killed. All hearts were then submitted for pathological evaluation.

RESULTS:

Acute PVI was successful in all 9/9 swine with lesion counts significantly lower in the SA arm (5.3 ± 5.9, 33.7 ± 10.0, and 28.0 ± 4.4 in SA, MN, and RF arms; p = .007 for SA and MN). At necropsy, circumferentiality and transmurality were 98% and 94% in SA, 98% and 80% in MN, and 100% and 100% in RF arms. A single steam pop was noted on sectioning in the SA arm swine and occurred in the high dose (18 W) strategy. Chronic PVI was acutely successful in 16/16 swine with no difference in PVI durability rates (62.5% vs. 75.0%), lesion transmurality (95.8 ± 17.4% vs.91.9 ± 25.9%), and circumferentiality (95.8 ± 6.6% vs. 94.8 ± 6.3%) between SA and MN arms. Catheter use time and lesion counts were lower in the SA arm compared to the MN arm (11.5 ± 12.7 vs. 21.8 ± 3.8 min, p = .046 and 4.8 ± 3.83 vs. 35.4 ± 4.4, p < .001).

CONCLUSION:

Motor-assisted semi-automated laser balloon ablation can improve upon procedural efficiency by reducing ablation time.
Asunto(s)
Palabras clave

Texto completo: 1 Colección: 01-internacional Banco de datos: MEDLINE Asunto principal: Venas Pulmonares / Fibrilación Atrial / Ablación por Catéter / Terapia por Láser Límite: Animals Idioma: En Revista: J Cardiovasc Electrophysiol Asunto de la revista: ANGIOLOGIA / CARDIOLOGIA / FISIOLOGIA Año: 2023 Tipo del documento: Article País de afiliación: Estados Unidos

Texto completo: 1 Colección: 01-internacional Banco de datos: MEDLINE Asunto principal: Venas Pulmonares / Fibrilación Atrial / Ablación por Catéter / Terapia por Láser Límite: Animals Idioma: En Revista: J Cardiovasc Electrophysiol Asunto de la revista: ANGIOLOGIA / CARDIOLOGIA / FISIOLOGIA Año: 2023 Tipo del documento: Article País de afiliación: Estados Unidos