Confounder-corrected T1 mapping in the liver through simultaneous estimation of T1 , PDFF, R 2 * , and B 1 + in a single breath-hold acquisition.
Roberts, Nathan T; Tamada, Daiki; Muslu, Yavuz; Hernando, Diego; Reeder, Scott B.
Afiliación
  • Roberts NT; Department of Radiology, University of Wisconsin - Madison, Madison, Wisconsin, USA.
  • Tamada D; Department of Electrical and Computer Engineering, University of Wisconsin - Madison, Madison, Wisconsin, USA.
  • Muslu Y; Department of Radiology, University of Wisconsin - Madison, Madison, Wisconsin, USA.
  • Hernando D; Department of Radiology, University of Wisconsin - Madison, Madison, Wisconsin, USA.
  • Reeder SB; Department of Biomedical Engineering, University of Wisconsin - Madison, Madison, Wisconsin, USA.
Magn Reson Med ; 89(6): 2186-2203, 2023 06.
Article en En | MEDLINE | ID: mdl-36656152
ABSTRACT

PURPOSE:

Quantitative volumetric T1 mapping in the liver has the potential to aid in the detection, diagnosis, and quantification of liver fibrosis, inflammation, and spatially resolved liver function. However, accurate measurement of hepatic T1 is confounded by the presence of fat and inhomogeneous B 1 + $$ {B}_1^{+} $$ excitation. Furthermore, scan time constraints related to respiratory motion require tradeoffs of reduced volumetric coverage and/or increased acquisition time. This work presents a novel 3D acquisition and estimation method for confounder-corrected T1 measurement over the entire liver within a single breath-hold through simultaneous estimation of T1 , fat and B 1 + $$ {B}_1^{+} $$ . THEORY AND

METHODS:

The proposed method combines chemical shift encoded MRI and variable flip angle MRI with a B 1 + $$ {B}_1^{+} $$ mapping technique to enable confounder-corrected T1 mapping. The method was evaluated theoretically and demonstrated in both phantom and in vivo acquisitions at 1.5 and 3.0T. At 1.5T, the method was evaluated both pre- and post- contrast enhancement in healthy volunteers.

RESULTS:

The proposed method demonstrated excellent linear agreement with reference inversion-recovery spin-echo based T1 in phantom acquisitions at both 1.5 and 3.0T, with minimal bias (5.2 and 45 ms, respectively) over T1 ranging from 200-1200 ms. In vivo results were in general agreement with reference saturation-recovery based 2D T1 maps (SMART1 Map, GE Healthcare).

CONCLUSION:

The proposed 3D T1 mapping method accounts for fat and B 1 + $$ {B}_1^{+} $$ confounders through simultaneous estimation of T1 , B 1 + $$ {B}_1^{+} $$ , PDFF and R 2 * $$ {R}_2^{\ast } $$ . It demonstrates strong linear agreement with reference T1 measurements, with low bias and high precision, and can achieve full liver coverage in a single breath-hold.
Asunto(s)
Palabras clave

Texto completo: 1 Colección: 01-internacional Banco de datos: MEDLINE Asunto principal: Enfermedad del Hígado Graso no Alcohólico / Hígado Límite: Humans Idioma: En Revista: Magn Reson Med Asunto de la revista: DIAGNOSTICO POR IMAGEM Año: 2023 Tipo del documento: Article País de afiliación: Estados Unidos

Texto completo: 1 Colección: 01-internacional Banco de datos: MEDLINE Asunto principal: Enfermedad del Hígado Graso no Alcohólico / Hígado Límite: Humans Idioma: En Revista: Magn Reson Med Asunto de la revista: DIAGNOSTICO POR IMAGEM Año: 2023 Tipo del documento: Article País de afiliación: Estados Unidos