Your browser doesn't support javascript.
loading
Implantable Defibrillator System Shock Function, Mortality, and Cause of Death After Magnetic Resonance Imaging.
Ra, Joshua; Oberdier, Matt T; Suzuki, Masahito; Vaidya, Dhananjay; Liu, Yisi; Hansford, Rozann; McVeigh, Diana; Weltin, Valeria; Tao, Susumu; Thiemann, David R; Nazarian, Saman; Halperin, Henry R.
Afiliación
  • Ra J; Department of Medicine, University of California San Francisco, San Francisco, California (J.R.).
  • Oberdier MT; Department of Cardiology, Johns Hopkins University, Baltimore, Maryland (M.T.O., M.S., D.V., R.H., D.M., V.W., S.T., D.R.T.).
  • Suzuki M; Department of Cardiology, Johns Hopkins University, Baltimore, Maryland (M.T.O., M.S., D.V., R.H., D.M., V.W., S.T., D.R.T.).
  • Vaidya D; Department of Cardiology, Johns Hopkins University, Baltimore, Maryland (M.T.O., M.S., D.V., R.H., D.M., V.W., S.T., D.R.T.).
  • Liu Y; Department of Pediatrics, Johns Hopkins University, Baltimore, Maryland (Y.L.).
  • Hansford R; Department of Cardiology, Johns Hopkins University, Baltimore, Maryland (M.T.O., M.S., D.V., R.H., D.M., V.W., S.T., D.R.T.).
  • McVeigh D; Department of Cardiology, Johns Hopkins University, Baltimore, Maryland (M.T.O., M.S., D.V., R.H., D.M., V.W., S.T., D.R.T.).
  • Weltin V; Department of Cardiology, Johns Hopkins University, Baltimore, Maryland (M.T.O., M.S., D.V., R.H., D.M., V.W., S.T., D.R.T.).
  • Tao S; Department of Cardiology, Johns Hopkins University, Baltimore, Maryland (M.T.O., M.S., D.V., R.H., D.M., V.W., S.T., D.R.T.).
  • Thiemann DR; Department of Cardiology, Johns Hopkins University, Baltimore, Maryland (M.T.O., M.S., D.V., R.H., D.M., V.W., S.T., D.R.T.).
  • Nazarian S; Department of Cardiology, University of Pennsylvania Perelman School of Medicine, Philadelphia, Pennsylvania (S.N.).
  • Halperin HR; Department of Cardiology and Departments of Radiology and Biomedical Engineering, Johns Hopkins University, Baltimore, Maryland (H.R.H.).
Ann Intern Med ; 176(3): 289-297, 2023 03.
Article en En | MEDLINE | ID: mdl-36716451
ABSTRACT

BACKGROUND:

Studies have shown that magnetic resonance imaging (MRI) does not have clinically important effects on the device parameters of non-MRI-conditional implantable cardioverter-defibrillators (ICDs). However, data on non-MRI-conditional ICD detection and treatment of arrhythmias after MRI are limited.

OBJECTIVE:

To examine if non-MRI-conditional ICDs have preserved shock function of arrhythmias after MRI.

DESIGN:

Prospective cohort study. (ClinicalTrials.gov NCT01130896).

SETTING:

1 center in the United States. PATIENTS 629 patients with non-MRI-conditional ICDs enrolled consecutively between February 2003 and January 2015.

INTERVENTIONS:

813 total MRI examinations at a magnetic field strength of 1.5 Tesla using a prespecified safety protocol. MEASUREMENTS Implantable cardioverter-defibrillator interrogations were collected after MRI. Clinical outcomes included arrhythmia detection and treatment, generator or lead exchanges, adverse events, and death.

RESULTS:

During a median follow-up of 2.2 years from MRI to latest available ICD interrogation before generator or lead exchange in 536 patients, 4177 arrhythmia episodes were detected, and 97 patients received ICD shocks. Sixty-one patients (10% of total) had 130 spontaneous ventricular tachycardia or fibrillation events terminated by ICD shocks. A total of 210 patients (33% of total) are known to have died (median, 1.7 years from MRI to death); 3 had cardiac arrhythmia deaths where shocks were indicated without direct evidence of device dysfunction.

LIMITATIONS:

Data were acquired at a single center and may not be generalizable to other clinical settings and MRI facilities. Implantable cardioverter-defibrillator interrogations were not available for a subset of patients; adjudication of cause of death relied solely on death certificate data in a subset.

CONCLUSION:

Non-MRI-conditional ICDs appropriately treated detected tachyarrhythmias after MRI. No serious adverse effects on device function were reported after MRI. PRIMARY FUNDING SOURCE Johns Hopkins University and National Institutes of Health.
Asunto(s)

Texto completo: 1 Colección: 01-internacional Banco de datos: MEDLINE Asunto principal: Desfibriladores Implantables Tipo de estudio: Etiology_studies / Guideline / Observational_studies Límite: Humans Idioma: En Revista: Ann Intern Med Año: 2023 Tipo del documento: Article

Texto completo: 1 Colección: 01-internacional Banco de datos: MEDLINE Asunto principal: Desfibriladores Implantables Tipo de estudio: Etiology_studies / Guideline / Observational_studies Límite: Humans Idioma: En Revista: Ann Intern Med Año: 2023 Tipo del documento: Article