A review of wear debris in thumb base joint implants.
Mangan, Fiona; Spece, Hannah; Weiss, Arnold-Peter C; Ladd, Amy L; Stockmans, Filip; Kurtz, Steven M.
Afiliación
  • Mangan F; Loci Orthopaedics, Galway, Ireland. fiona@lociorthopaedics.com.
  • Spece H; Drexel University, Philadelphia, PA, USA.
  • Weiss AC; Gyroid, LLC, Haddonfield, NJ, USA.
  • Ladd AL; Brown University, Providence, RI, USA.
  • Stockmans F; Stanford University, Stanford, CA, USA.
  • Kurtz SM; KU Leuven, Louvain, Belgium.
Eur J Orthop Surg Traumatol ; 34(1): 251-269, 2024 Jan.
Article en En | MEDLINE | ID: mdl-37439887
ABSTRACT

AIM:

Polymers and metals, such as polyethylene (PE) and cobalt chrome (CoCr), are common materials used in thumb-based joint implants, also known as CMC (Carpometacarpal) arthroplasty. The purpose of this review was to investigate the reported failure modes related to wear debris from these type of materials in CMC implants. The impact of wear debris on clinical outcomes of CMC implants was also examined. Potential adverse wear conditions and inflammatory particle characteristics were also considered.

METHOD:

A literature search was performed using PRISMA guidelines and 55 studies were reviewed including 49 cohort studies and 6 case studies. Of the 55 studies, 38/55 (69%) focused on metal-on-polyethylene devices, followed by metal-on-metal (35%), and metal-on-bone (4%).

RESULTS:

The summarized data was used to determine the frequency of failure modes potentially related to wear debris from metals and/or polymers. The most commonly reported incidents potentially relating to debris were implant loosening (7.1%), osteolysis (1.2%) and metallosis (0.6%). Interestingly the reported mechanisms behind osteolysis and loosening greatly varied. Inflammatory reactions, while rare, were generally attributed to metallic debris from metal-on-metal devices. Mechanisms of adverse wear conditions included implant malpositioning, over-tensioning, high loading for active patients, third-body debris, and polyethylene wear-through. No specific examination of debris particle characterization was found, pointing to a gap in the literature.

CONCLUSION:

This review underscores the types of failure modes associated with wear debris in CMC implants. It was found that failure rates and adverse wear conditions of CMC implants of any design are low and the exact relationship between wear debris and implant incidences, such as osteolysis and loosening remains uncertain. The authors note that further research and specific characterization is required to understand the relationship between debris and implant failure.
Asunto(s)
Palabras clave

Texto completo: 1 Colección: 01-internacional Banco de datos: MEDLINE Asunto principal: Osteólisis Tipo de estudio: Guideline / Observational_studies / Risk_factors_studies Límite: Humans Idioma: En Revista: Eur J Orthop Surg Traumatol Año: 2024 Tipo del documento: Article País de afiliación: Irlanda

Texto completo: 1 Colección: 01-internacional Banco de datos: MEDLINE Asunto principal: Osteólisis Tipo de estudio: Guideline / Observational_studies / Risk_factors_studies Límite: Humans Idioma: En Revista: Eur J Orthop Surg Traumatol Año: 2024 Tipo del documento: Article País de afiliación: Irlanda