Minimize Precolumn Band Broadening with Immiscible Solvent Sandwich Injection.
Zhu, Koudi; Pursch, Matthias; Huygens, Bram; Eeltink, Sebastiaan; Desmet, Gert.
Afiliación
  • Zhu K; IFF, Pharma Solutions, 1801 Larkin Center Drive, Midland, Michigan 48640, United States.
  • Pursch M; Department of Chemical Engineering, Vrije Universiteit Brussel, Pleinlaan 2, 1050 Brussel, Belgium.
  • Huygens B; Dow Deutschland Anlagen GmbH, Analytical Science, 65201 Wiesbaden, Germany.
  • Eeltink S; Department of Chemical Engineering, Vrije Universiteit Brussel, Pleinlaan 2, 1050 Brussel, Belgium.
  • Desmet G; Department of Chemical Engineering, Vrije Universiteit Brussel, Pleinlaan 2, 1050 Brussel, Belgium.
Anal Chem ; 95(41): 15311-15317, 2023 Oct 17.
Article en En | MEDLINE | ID: mdl-37797306
ABSTRACT
We investigated the possibility of reducing the effect of precolumn band broadening (PreCBB) by sandwiching the sample between two small plugs of an immiscible liquid. It has been found that in cases of severe PreCBB, improvements in peak efficiency can amount up to 20 times for the early-eluting compounds. For smaller degrees of PreCBB, the gain on the efficiency of early-eluting compounds is smaller (order of 50%), yet it is still significant. It has been verified that the presence of the immiscible fluid sandwich does not affect the repeatability of the analysis nor the linearity of the calibration curves used for analyte quantitation. It is also shown that the main effect of the two sandwich plugs is the minimization of the dispersion in the precolumn transfer tubing itself, which makes the method fundamentally different from pure on-column focusing methods such as the performance optimizing injection sequence (POISe) method. It is further demonstrated that both halves of the sandwich are needed, since the beneficial effect is clearly much smaller when only one plug is present. A drawback of the method is that some of the late-eluting peaks are slightly adversely affected by the presence of the sandwich liquid in the case where 127 µm i.d. tubing was used. The mechanism for this peak deterioration effect is at present still unclear but only occurs under gradient conditions and is clearly linked to the size of the sandwich plugs (the smaller the plugs, the smaller the adverse effect) and the internal diameter of the tubing used between the injection valve and the column.

Texto completo: 1 Colección: 01-internacional Banco de datos: MEDLINE Idioma: En Revista: Anal Chem Año: 2023 Tipo del documento: Article País de afiliación: Estados Unidos

Texto completo: 1 Colección: 01-internacional Banco de datos: MEDLINE Idioma: En Revista: Anal Chem Año: 2023 Tipo del documento: Article País de afiliación: Estados Unidos