In-vitro validation of coronary physiology assessment with 5 French guiding catheters.
Benedetti, Alice; Castaldi, Gianluca; Poletti, Enrico; Moroni, Alice; El Jattari, Hicham; Scott, Benjamin; Convens, Carl; Verheye, Stefan; Vermeersch, Paul; Agostoni, Pierfrancesco; Zivelonghi, Carlo.
Afiliación
  • Benedetti A; HartCentrum, Ziekenhuis Netwerk Antwerpen (ZNA) Middelheim, Antwerp, Belgium.
  • Castaldi G; HartCentrum, Ziekenhuis Netwerk Antwerpen (ZNA) Middelheim, Antwerp, Belgium.
  • Poletti E; HartCentrum, Ziekenhuis Netwerk Antwerpen (ZNA) Middelheim, Antwerp, Belgium; Clinical and Interventional Cardiology Department, IRCCS Policlinico San Donato, San Donato Milanese, Italy.
  • Moroni A; HartCentrum, Ziekenhuis Netwerk Antwerpen (ZNA) Middelheim, Antwerp, Belgium.
  • El Jattari H; HartCentrum, Ziekenhuis Netwerk Antwerpen (ZNA) Middelheim, Antwerp, Belgium.
  • Scott B; HartCentrum, Ziekenhuis Netwerk Antwerpen (ZNA) Middelheim, Antwerp, Belgium.
  • Convens C; HartCentrum, Ziekenhuis Netwerk Antwerpen (ZNA) Middelheim, Antwerp, Belgium.
  • Verheye S; HartCentrum, Ziekenhuis Netwerk Antwerpen (ZNA) Middelheim, Antwerp, Belgium.
  • Vermeersch P; HartCentrum, Ziekenhuis Netwerk Antwerpen (ZNA) Middelheim, Antwerp, Belgium.
  • Agostoni P; HartCentrum, Ziekenhuis Netwerk Antwerpen (ZNA) Middelheim, Antwerp, Belgium.
  • Zivelonghi C; HartCentrum, Ziekenhuis Netwerk Antwerpen (ZNA) Middelheim, Antwerp, Belgium. Electronic address: carlo.zivelonghi@gmail.com.
Cardiovasc Revasc Med ; 60: 55-63, 2024 Mar.
Article en En | MEDLINE | ID: mdl-37863759
ABSTRACT

BACKGROUND:

The trans-radial approach for cardiac catheterization led to an increasing adoption of 5 French (F) catheters. We aim to evaluate reliability and reproducibility of coronary physiology assessment performed with 5F guiding catheter (GC).

METHODS:

Physiological measurements were performed in a coronary flow simulator, which provides two pulsatile flows, the baseline and hyperaemic flows. Two screws, positioned proximally and distally to the distal sensor of a pressure-temperature guidewire, were used to determine various combinations of stenoses and distal obstructions, simulating different pathophysiological conditions. For each setting, 5 measurements of fractional flow reserve (FFR), coronary flow reserve (CFR) and index of microvascular resistance (IMR) were performed with 6F and 5F GCs.

RESULTS:

A total amount of 190 measurements were performed, 95 with 6F GC and 95 with 5F GC. Minimal differences between 6F and 5F GCs were detected for FFR [0.91 (IQR 0.87-0.94) and 0.87 (IQR 0.82-0.92) respectively, p < 0.001] and IMR (16.5 ± 8.8 and 15.4 ± 8.3 respectively, p = 0.001). Mean CFR was comparable between 6F and 5F GCs (3.6 ± 1.1 and 3.5 ± 0.7 respectively, p = 0.38). Misclassification rates were 1.0 %, 1.0 % and 0 % for FFR, CFR and IMR, respectively. According to Passing-Bablok analysis, an excellent agreement between 6F and 5F GCs was demonstrated for FFR and IMR, and a modest agreement for CFR. All measurements with 5F GC showed high reproducibility.

CONCLUSIONS:

In our in-vitro model, a complete physiological assessment including FFR, CFR and IMR resulted substantially comparable between 6F and 5F GCs. Further in-vivo analysis is required to support these findings.
Asunto(s)
Palabras clave

Texto completo: 1 Colección: 01-internacional Banco de datos: MEDLINE Asunto principal: Estenosis Coronaria / Reserva del Flujo Fraccional Miocárdico Límite: Humans Idioma: En Revista: Cardiovasc Revasc Med Asunto de la revista: ANGIOLOGIA / CARDIOLOGIA Año: 2024 Tipo del documento: Article País de afiliación: Bélgica

Texto completo: 1 Colección: 01-internacional Banco de datos: MEDLINE Asunto principal: Estenosis Coronaria / Reserva del Flujo Fraccional Miocárdico Límite: Humans Idioma: En Revista: Cardiovasc Revasc Med Asunto de la revista: ANGIOLOGIA / CARDIOLOGIA Año: 2024 Tipo del documento: Article País de afiliación: Bélgica