Accuracy and User Acceptability of 24-hour Ambulatory Blood Pressure Monitoring by a Prototype Cuffless Multi-Sensor Device Compared to a Conventional Oscillometric Device.
Heimark, Sondre; Hove, Christine; Stepanov, Alexey; Boysen, Elin Sundby; Gløersen, Øyvind; Bøtke-Rasmussen, Kasper Gade; Gravdal, Hans Jacob; Narayanapillai, Kesi; Fadl Elmula, Fadl Elmula M; Seeberg, Trine M; Larstorp, Anne Cecilie K; Waldum-Grevbo, Bård.
Afiliación
  • Heimark S; Department of Nephrology, Oslo University Hospital, Ullevål, Oslo, Norway.
  • Hove C; Institute of Clinical Medicine, University of Oslo, Oslo, Norway.
  • Stepanov A; Department of Nephrology, Oslo University Hospital, Ullevål, Oslo, Norway.
  • Boysen ES; Institute of Clinical Medicine, University of Oslo, Oslo, Norway.
  • Gløersen Ø; Aidee Health AS, Oslo, Norway.
  • Bøtke-Rasmussen KG; Department of Smart Sensors and Microsystems, SINTEF Digital, Oslo, Norway.
  • Gravdal HJ; Department of Smart Sensors and Microsystems, SINTEF Digital, Oslo, Norway.
  • Narayanapillai K; Aidee Health AS, Oslo, Norway.
  • Fadl Elmula FEM; Department of Smart Sensors and Microsystems, SINTEF Digital, Oslo, Norway.
  • Seeberg TM; Sandefjord Helsepark, Sandefjord, Norway.
  • Larstorp ACK; Department of Smart Sensors and Microsystems, SINTEF Digital, Oslo, Norway.
  • Waldum-Grevbo B; Section for Cardiovascular and Renal Research, Oslo University Hospital, Ullevål, Oslo, Norway.
Blood Press ; 32(1): 2274595, 2023 12.
Article en En | MEDLINE | ID: mdl-37885101
What is the context?Hypertension is a major risk factor for cardiovascular and cerebrovascular end-organ damage, morbidity, and mortality world-wide.Accurate measurement of blood pressure is essential for the diagnosis and management of hypertension.What is new?Cuffless blood pressure devices that allow measurement of blood pressure without a pressure cuff is a promising and novel method of blood pressure estimation.The objective of this study is to assess whether pulse arrival time alone can be used to estimate blood pressure accurately during 24-hour ambulatory blood pressure monitoring, using a prototype cuffless device placed on the chest.Our analysis shows that a general model based on pulse arrival time overestimated ambulatory blood pressure, especially during nighttime.User acceptability was higher with the cuffless device compared to a conventional cuff-based oscillometric device during 24-hour ambulatory blood pressure monitoring.What is the impact?This study provides further evidence that accurate blood pressure estimations cannot be achieved by using pulse arrival time alone as a surrogate for blood pressure measurements.
Asunto(s)
Palabras clave

Texto completo: 1 Colección: 01-internacional Banco de datos: MEDLINE Asunto principal: Monitoreo Ambulatorio de la Presión Arterial / Hipertensión Límite: Adult / Humans Idioma: En Revista: Blood Press Asunto de la revista: ANGIOLOGIA Año: 2023 Tipo del documento: Article País de afiliación: Noruega

Texto completo: 1 Colección: 01-internacional Banco de datos: MEDLINE Asunto principal: Monitoreo Ambulatorio de la Presión Arterial / Hipertensión Límite: Adult / Humans Idioma: En Revista: Blood Press Asunto de la revista: ANGIOLOGIA Año: 2023 Tipo del documento: Article País de afiliación: Noruega