Sonorheometry Device Thresholds in Liver Transplantation: An Observational Retrospective Study.
Soucy-Proulx, Maxim; Kato, Hiromi; Coeckelenbergh, Sean; Naili Kortaia, Salima; Herboulier, Laurence; Pittau, Gabriella; Pham, Patrick; Lemoine, Antoinette; Duranteau, Jacques; Roullet, Stéphanie.
Afiliación
  • Soucy-Proulx M; Service d'Anesthésie-Réanimation, Hôpital Paul Brousse, Assistance Publique-Hôpitaux de Paris (AP-HP), Université Paris Saclay, 94800 Villejuif, France.
  • Kato H; Department of Anesthesiology, Centre Hospitalier de l'Université de Montréal, Montreal, QC H2X 0A9, Canada.
  • Coeckelenbergh S; Service d'Anesthésie-Réanimation, Hôpital Paul Brousse, Assistance Publique-Hôpitaux de Paris (AP-HP), Université Paris Saclay, 94800 Villejuif, France.
  • Naili Kortaia S; Service d'Anesthésie-Réanimation, Hôpital Paul Brousse, Assistance Publique-Hôpitaux de Paris (AP-HP), Université Paris Saclay, 94800 Villejuif, France.
  • Herboulier L; Outcomes Research Consortium, Cleveland, OH 44195, USA.
  • Pittau G; Service d'Anesthésie-Réanimation, Hôpital Paul Brousse, Assistance Publique-Hôpitaux de Paris (AP-HP), Université Paris Saclay, 94800 Villejuif, France.
  • Pham P; Service d'Anesthésie-Réanimation, Hôpital Paul Brousse, Assistance Publique-Hôpitaux de Paris (AP-HP), Université Paris Saclay, 94800 Villejuif, France.
  • Lemoine A; Centre Hépato-Biliaire, Hôpital Paul Brousse, Assistance Publique-Hôpitaux de Paris (AP-HP), Université Paris Saclay, 94800 Villejuif, France.
  • Duranteau J; Service de Biochimie et Oncogénétique, Hôpital Paul Brousse, Assistance Publique-Hôpitaux de Paris (AP-HP), Université Paris Saclay, 94800 Villejuif, France.
  • Roullet S; Service de Biochimie et Oncogénétique, Hôpital Paul Brousse, Assistance Publique-Hôpitaux de Paris (AP-HP), Université Paris Saclay, 94800 Villejuif, France.
J Clin Med ; 13(3)2024 Jan 25.
Article en En | MEDLINE | ID: mdl-38337389
ABSTRACT

BACKGROUND:

Liver transplantation (LT) remains a potentially haemorrhagic procedure whose perioperative bleeding and transfusion could be better monitored using point-of-care devices. Quantra® is a device based on sonorheometry to assess whole blood clot formation. Our aims were to describe Quantra® parameters during LT and to study their correlations with standard laboratory parameters, and to determine Quantra® cut-off values for thrombocytopenia, hypofibrinogenemia and coagulation factors' deficit.

METHODS:

In 34 patients undergoing LT, blood samples were collected before surgical incision, 15 min after the beginning of the anhepatic phase, and 15 min after arterial revascularization of the graft.

RESULTS:

Clotting time (CT) was well correlated with prothrombin (PT) ratio and activated partial thromboplastin time (aPTT) ratio. Platelet contribution to clot stiffness (PCS) was correlated with platelets (ρ = 0.82, p < 0.001) and fibrinogen contribution clot stiffness (FCS) with fibrinogen (Fg) (ρ = 0.74, p < 0.001). CT predicted a PT ratio < 30% with an area under the curve (AUC) of 0.93 (95% CI 0.87-0.98; p < 0.001). PCS predicted a platelet count < 50 G/L with an AUC of 0.87 (95% CI 0.76-0.98, p < 0.001). FCS predicted a Fg < 1.0, 1.2 or 1.5 g/L, with an AUC of 0.86 (95% CI 0.77-094, p < 0.001), 0.82 (95% CI 0.74-0.91, p < 0.001) and 0.88 (95% CI 0.82-0.95, p < 0.001), respectively.

CONCLUSION:

Quantra® provides a rapid assessment of haemostasis during LT.