Design and experiment of high-field asymmetric waveform ion mobility spectrometry chip based on an integrated temperature control printed circuit board structure.
Zeng, Hao; Zhang, Yu-Qiao; Wang, Yi-Fei; Du, Xiao-Xia; Xiao, Wen-Xiang; Li, Hua.
Afiliación
  • Zeng H; School of Life and Environmental Sciences, Guilin University of Electronic Technology, Guilin, Guangxi, China.
  • Zhang YQ; School of Life and Environmental Sciences, Guilin University of Electronic Technology, Guilin, Guangxi, China.
  • Wang YF; School of Life and Environmental Sciences, Guilin University of Electronic Technology, Guilin, Guangxi, China.
  • Du XX; School of Life and Environmental Sciences, Guilin University of Electronic Technology, Guilin, Guangxi, China.
  • Xiao WX; Guangxi Colleges and Universities Key Laboratory of Biomedical Sensors and Intelligent Instruments, Guilin, Guangxi, China.
  • Li H; School of Life and Environmental Sciences, Guilin University of Electronic Technology, Guilin, Guangxi, China.
Rapid Commun Mass Spectrom ; 38(5): e9699, 2024 Mar 15.
Article en En | MEDLINE | ID: mdl-38355881
ABSTRACT
RATIONALE During the detection of volatile organic compounds (VOC) using high-field asymmetric waveform ion mobility spectrometry (FAIMS), the ambient temperature significantly impacts the accuracy of planar FAIMS. To mitigate the influence of ambient temperature on detection accuracy and enhance resolution, a FAIMS system based on the inner impedance characteristics of a printed circuit board (PCB) was designed for temperature control.

METHODS:

This study, conducted under standard atmospheric pressure, aimed to assess the signal stability of a planar FAIMS instrument with and without temperature control, and the effect of temperature change on the detection ability of acetone, ethanol, and their mixture was studied using PCB self-heating.

RESULTS:

Experimental results demonstrated that the base noise in FAIMS with temperature control was 0.2 pA, whereas that in FAIMS without temperature control was 1.8 pA. Notably, with increasing temperature, the detection ability of FAIMS changes accordingly. The optimal relative detection ability of acetone was observed when the electrode plate was heated to 45°C under an electric field of 15 kV/cm.

CONCLUSIONS:

This study provides a novel approach to improve the resolving power of FAIMS systems and their signal-to-noise ratio. The utilization of a PCB-based temperature control proved effective in stabilizing FAIMS signal characteristics and optimizing detection capabilities, particularly for VOCs such as acetone. These findings have significant implications for improving the accuracy and resolving power of FAIMS systems in VOC detection applications.

Texto completo: 1 Colección: 01-internacional Banco de datos: MEDLINE Idioma: En Revista: Rapid Commun Mass Spectrom Año: 2024 Tipo del documento: Article País de afiliación: China