Rapid Detection of the Anti-Tumor Drug Etoposide in Biological Samples by Using a Nanoporous-Gold-Based Electrochemical Sensor.
Yu, Huiyuan; Hu, Mengjie; Wang, Xiaolei; Wang, Xia; Xun, Luying; Liu, Honglei.
Afiliación
  • Yu H; State Key Laboratory of Microbial Technology, Shandong University, 72 Binhai Road, Qingdao 266237, China.
  • Hu M; State Key Laboratory of Microbial Technology, Shandong University, 72 Binhai Road, Qingdao 266237, China.
  • Wang X; State Key Laboratory of Microbial Technology, Shandong University, 72 Binhai Road, Qingdao 266237, China.
  • Wang X; State Key Laboratory of Microbial Technology, Shandong University, 72 Binhai Road, Qingdao 266237, China.
  • Xun L; State Key Laboratory of Microbial Technology, Shandong University, 72 Binhai Road, Qingdao 266237, China.
  • Liu H; School of Molecular Biosciences, Washington State University, Pullman, WA 99164-7520, USA.
Molecules ; 29(5)2024 Feb 28.
Article en En | MEDLINE | ID: mdl-38474572
ABSTRACT
Monitoring etoposide is important due to its wide usage in anti-tumor therapy; however, the commonly used HPLC method is expensive and often requires complicated extraction and detection procedures. Electrochemical analysis has great application prospects because of its rapid response and high specificity, sensitivity, and efficiency with low cost and high convenience. In this study, we constructed a nanoporous gold (NPG)-modified GCE for the detection of etoposide. The electrochemical oxidation of etoposide by NPG caused a sensitive current peak at +0.27 V with good reproductivity in 50 mM of phosphate buffer (pH 7.4). The relationship between etoposide concentration and peak current was linear in the range between 0.1 and 20 µM and between 20 and 150 µM, with a detection sensitivity of 681.8 µA mM-1 cm-2 and 197.2 µA mM-1 cm-2, respectively, and a limit of detection (LOD) reaching 20 nM. The electrode had a good anti-interference ability to several common anions and cations. Spiked recovery tests in serum, urine, and fermentation broth verified the excellent performance of the sensor in terms of sensitivity, reproducibility, and specificity. This may provide a promising tool for the detection of etoposide in biological samples.