Your browser doesn't support javascript.
loading
[Direct Positron Emission Imaging Using Ultrafast Timing Performance Detectors].
Ota, Ryosuke; Kwon, Sun Il; Hashimoto, Fumio; Nakajima, Kyohei; Ogawa, Izumi; Tamagawa, Yoichi; Omura, Tomohide; Hasegawa, Tomoyuki; Cherry, Simon R.
Afiliación
  • Ota R; Central Research Laboratory, Hamamatsu Photonics K.K.
  • Kwon SI; Department of Biomedical Engineering, University of California, Davis, USA.
  • Eric Berg; Department of Biomedical Engineering, University of California, Davis, USA.
  • Hashimoto F; Central Research Laboratory, Hamamatsu Photonics K.K.
  • Nakajima K; School of Engineering, University of Fukui.
  • Ogawa I; School of Engineering, University of Fukui.
  • Tamagawa Y; School of Engineering, University of Fukui.
  • Omura T; Central Research Laboratory, Hamamatsu Photonics K.K.
  • Hasegawa T; School of Allied Health Sciences, Kitasato University.
  • Cherry SR; Department of Biomedical Engineering, University of California, Davis, USA.
Igaku Butsuri ; 44(2): 29-35, 2024.
Article en Ja | MEDLINE | ID: mdl-38945880
ABSTRACT
This is an explanatory paper on Sun Il Kwon et al., Nat. Photon. 15 914-918, 2021 and some parts of this manuscript are translated from the paper. Medical imaging modalities such as X-ray computed tomography, Magnetic resonance imaging, positron emission tomography (PET), and single photon emission computed tomography, require image reconstruction processes, consequently constraining them to form cylindrical shapes. However, among them, only PET can use additional information, so called time of flight, on an event-by-event basis. If coincidence time resolution (CTR) of PET detectors improved to 30 ps, which corresponds to spatial resolution of 4.5 mm, directly localizing electron-positron annihilation point is possible, allowing us to circumvent image reconstruction processes and free us from the geometric constraint. We call this concept direct positron emission imaging (dPEI). We have developed ultrafast radiation detectors by focusing on Cherenkov photon detection. Furthermore, the CTR of 32 ps being equivalent to 4.8 mm spatial resolution is achieved by combining deep learning-based signal processing with the detectors. In this article, we explain how we developed the detectors and demonstrated the first dPEI using different types of phantoms, how we will tackle limitations to be addressed to make the dPEI more practical, and how dPEI will emerge as an imaging modality in nuclear medicine.
Asunto(s)
Palabras clave

Texto completo: 1 Colección: 01-internacional Banco de datos: MEDLINE Asunto principal: Procesamiento de Imagen Asistido por Computador / Tomografía de Emisión de Positrones Idioma: Ja Revista: Igaku Butsuri Asunto de la revista: BIOFISICA Año: 2024 Tipo del documento: Article

Texto completo: 1 Colección: 01-internacional Banco de datos: MEDLINE Asunto principal: Procesamiento de Imagen Asistido por Computador / Tomografía de Emisión de Positrones Idioma: Ja Revista: Igaku Butsuri Asunto de la revista: BIOFISICA Año: 2024 Tipo del documento: Article