Your browser doesn't support javascript.
loading
Performance evaluation of the 3D-ring cadmium-zinc-telluride (CZT) StarGuide system according to the NEMA NU 1-2018 standard.
Zorz, Alessandra; Rossato, Marco Andrea; Turco, Paolo; Colombo Gomez, Luca Maria; Bettinelli, Andrea; De Monte, Francesca; Paiusco, Marta; Zucchetta, Pietro; Cecchin, Diego.
Afiliación
  • Zorz A; Medical Physics Department, Veneto Institute of Oncology IOV - IRCCS, Padua, Italy. alessandra.zorz@iov.veneto.it.
  • Rossato MA; Medical Physics Department, Veneto Institute of Oncology IOV - IRCCS, Padua, Italy.
  • Turco P; Unit of Nuclear Medicine, University Hospital of Padova, Padua, Italy.
  • Colombo Gomez LM; Medical Physics Department, Veneto Institute of Oncology IOV - IRCCS, Padua, Italy.
  • Bettinelli A; Medical Physics Department, Veneto Institute of Oncology IOV - IRCCS, Padua, Italy.
  • De Monte F; Medical Physics Department, Veneto Institute of Oncology IOV - IRCCS, Padua, Italy.
  • Paiusco M; Medical Physics Department, Veneto Institute of Oncology IOV - IRCCS, Padua, Italy.
  • Zucchetta P; Unit of Nuclear Medicine, University Hospital of Padova, Padua, Italy.
  • Cecchin D; Unit of Nuclear Medicine, Department of Medicine (DIMED), University Hospital of Padova, Padua, Italy.
EJNMMI Phys ; 11(1): 69, 2024 Jul 25.
Article en En | MEDLINE | ID: mdl-39052176
ABSTRACT

BACKGROUND:

The application of semi-conductor detectors such as cadmium-zinc-telluride (CZT) in nuclear medicine improves extrinsic energy resolution and count sensitivity due to the direct conversion of gamma photons into electric signals. A 3D-ring pixelated CZT system named StarGuide was recently developed and implemented by GE HealthCare for SPECT acquisition. The system consists of 12 detector columns with seven modules of 16 × 16 CZT pixelated crystals, each with an integrated parallel-hole tungsten collimator. The axial coverage is 27.5 cm. The detector thickness is 7.25 mm, which allows acquisitions in the energy range [40-279] keV. Since there is currently no performance characterization specific to 3D-ring CZT SPECT systems, the National Electrical Manufacturers Association (NEMA) NU 1-2018 clinical standard can be tailored to these cameras. The aim of this study was to evaluate the performance of the SPECT/CT StarGuide system according to the NEMA NU 1-2018 clinical standard specifically adapted to characterize the new 3D-ring CZT.

RESULTS:

Due to the integrated collimator, the system geometry and the pixelated nature of the detector, some NEMA tests have been adapted to the features of the system. The extrinsic measured energy resolution was about 5-6% for the tested isotopes (99mTc, 123I and 57Co); the maximum count rate was 760 kcps and the observed count rate at 20% loss was 917 kcps. The system spatial resolution in air extrapolated at 10 cm with 99mTc was 7.2 mm, while the SPECT spatial resolutions with scatter were 4.2, 3.7 and 3.6 mm in a central, radial and tangential direction respectively. Single head sensitivity value for 99mTc was 97 cps/MBq; with 12 detector columns, the system volumetric sensitivity reached 520 kcps MBq-1 cc-1.

CONCLUSIONS:

The performance tests of the StarGuide can be performed according to the NEMA NU 1-2018 standard with some adaptations. The system has shown promising results, particularly in terms of energy resolution, spatial resolution and volumetric sensitivity, potentially leading to higher quality clinical images.
Palabras clave

Texto completo: 1 Colección: 01-internacional Banco de datos: MEDLINE Idioma: En Revista: EJNMMI Phys Año: 2024 Tipo del documento: Article País de afiliación: Italia

Texto completo: 1 Colección: 01-internacional Banco de datos: MEDLINE Idioma: En Revista: EJNMMI Phys Año: 2024 Tipo del documento: Article País de afiliación: Italia